VitalThings Guardian Contactless Monitoring

NCT06083272 — Completed, 41 participants, VitalThings

ClinicalTrials Daily  ·  Areas  ·  Conditions  ·  Drugs  ·  Sponsors  ·  
RegistryNCT06083272, acronym VINCENT
StatusCompleted
Study typeObservational
Enrollment41 (Actual)
SponsorVitalThings (Industry); collaborators: St. Olavs Hospital
ConditionsRespiratory Depression, Respiratory Failure, Deterioration, Clinical
InterventionsVitalthings Guardian M10 fixed and M10 mobile (Device)
Datesstart 2023-11-15, primary completion 2023-11-28, completion 2023-11-28, first posted 2023-10-13, last updated 2023-12-05
Locations1 site in Norway

Related Papers Related Patents Related Grants Related Experts
All trials for Respiratory Depression All trials for Respiratory Failure All trials of Vitalthings Guardian M10 fixed and M10 mobile

-Summary

This is a confirmatory study without any intervention. It is an uncontrolled, non-randomized and open-label study with measurements made with comparators, and it has a preset hypothesis for the primary endpoint. There are no similar devices to VitalThings Guardian M10 / M10 mobile on the market, consequently one or more different types of devices must be used as comparators.

Expand all sections  ·  Collapse all

+Detailed description

The purpose of the clinician investigation is to validate a claim made for the investigational device performance. The objective is to confirm that the device performs at +/- 1.2 respirations per minute (RPM) accuracy against the identified comparator device, and compare performance to the standard of care as documented in the literature. The study will be conducted with patients admitted to the emergency ward after the initial treatment and examination.

+Arms and interventions

ArmTypeInterventionsDescription
Study populationDevice: Vitalthings Guardian M10 fixed and M10 mobileThe study population is comprised of two subgroups; 1. Adult patients admitted to the Emergency Ward (EW). Patients are recruited after the treatment and examination in EW are finished and the patient is waiting for transfer to a hospital ward. 2. If there is a prolonged period of waiting time for a control blood sample, chest CT or any other examination which usually incurs significant waiting time in the EW. This will be evaluated by a specialist in emergency medicine based on the premise that the inclusion in the study will not delay or interfere with these examinations.
  • Vitalthings Guardian M10 fixed and M10 mobile (Device), also known as Nox T3s (NOX Medical, Alpharetta, GA, USA ) with a Nonin 3150 pulse oximeter (Nonin Medical Inc., MN, USA): VitalThings Guardian M10 / M10 mobile (VitalThings, Norway), are the investigational devices. These are the same technical devices made in two versions: one wall-mounted (VitalThings Guardian M10 fixed) and one mounted on a trolley (VitalThings Guardian M10 mobile). The Nox T3s and Nonin will be used as comparator devices.

+Outcome measures

Primary outcomes

  • Accuracy [Will be assessed after last participant has finished the investigation. One interim analysis in first week is planned.]: Documented system respiration frequency in RPM for both investigational device and comparator.

Secondary outcomes

  • Comparative pulse data [Will be assessed in 2024, furter details not established.]: Gather comparative pulse data to be used for retrospective validation of algorithm performance for contactless heart rate measurement/monitoring.

Other outcomes

  • Acceptance criteria respiratory rate [Will be assessed after last participant has finished the investigation. One interim analysis in first week is planned.]: A limit of agreement of ±1.2 RPM against comparators in setting the respiration rate.

+Eligibility

Sex: All, ages: 18 Years to any, accepts healthy volunteers

Inclusion Criteria:

* Adult patients admitted to the Emergency Ward (EW)

Exclusion Criteria:

* Under age (below 18 years) and subjects not able to provide informed consent themselves. Subjects dismissed from the EW rooms within 30 minutes from first registration are removed from the study.

+Locations (1)

FacilityCityCountryStatus
St Olavs University HospitalTrondheimNorway

+Publications

  • Results publication: Toften S, Kjellstadli JT, Kvaerness J, Pedersen L, Laugsand LE, Thu OKF. Contactless and continuous monitoring of respiratory rate in a hospital ward: a clinical validation study. Front Physiol. 2024 Nov 27;15:1502413. doi: 10.3389/fphys.2024.1502413. eCollection 2024. PMID 39665054