Evaluation of AutoSense OS on a Cochlear Implant Sound Processor Programmed With Target CI Fitting Software in Adult Users of the HiResolution Bionic Ear System
| Registry | NCT04237207 |
|---|---|
| Status | Completed |
| Phase | Not applicable |
| Study type | Interventional; non randomized, parallel, masking: none, treatment |
| Enrollment | 22 (Actual) |
| Sponsor | Advanced Bionics (Industry) |
| Conditions | Hearing Loss, Ear Diseases, Hearing Disorders, Otorhinolaryngologic Diseases |
| Interventions | Naída Cochlear Implant Q90 sound processor (Naida CI Q90) & software, 301-M062 sound processor & software (Device) |
| Dates | start 2019-12-17, primary completion 2020-10-15, completion 2020-10-15, first posted 2020-01-23, last updated 2022-05-13 |
| Locations | 6 sites in United States |
Related Papers Related Patents Related Grants Related Experts
All trials for Hearing Loss All trials for Ear Diseases All trials of Naída Cochlear Implant Q90 sound processor (Naida CI Q90) & software All trials of 301-M062 sound processor & software
This is a prospective within-subjects repeated-measures study that will enroll 24 adult users implanted with a HiResolution Bionic Ear System (HiRes 90K or newer). Subjects will be assigned to one of two cohorts based on audiometric results at the Baseline visit
The overall goal of this clinical study is to demonstrate the safety and efficacy of a cochlear implant sound processor. Hearing outcomes are expected to be similar non-inferior to as compared to the currently approved software and processors. Therefore, the study described herein uses a non-inferiority design to determine whether sentence recognition in quiet and in noise is no worse with the new sound processor than with currently approved software on a Q90 processor.
| Arm | Type | Interventions | Description |
|---|---|---|---|
| Aidable Residual Hearing (ARH) Cohort | Other | Device: Naída Cochlear Implant Q90 sound processor (Naida CI Q90) & software; Device: 301-M062 sound processor & software | Control Device followed by experimental Device. |
| Electric Only (EO) Cohort | Other | Device: Naída Cochlear Implant Q90 sound processor (Naida CI Q90) & software; Device: 301-M062 sound processor & software | Control Device followed by experimental Device. |
Sex: All, ages: 18 Years to any
General Inclusion Criteria: * Ability to provide Informed Consent * 18 years of age or older * Unilateral user of a HiResolutionTM Bionic Ear System (HiRes 90KTM, HiRes 90KTM Advantage, HiResTM Ultra, HiResTM Ultra 3D), including bilaterally implanted subjects that only use one implant in their everyday listening modality * Minimum of 6 months of CI experience * Having used a Naída CI Q70 or a Naída CI Q90 as their primary processor for a minimum of one month * Presently using a current steering strategy * At least moderate open-set speech recognition abilities with implant alone, as defined by achieving a score of ≥ 60% words correct in the AzBio in quiet test using the Naída CI Q90 research processor * English language proficiency as determined by the investigator * Willingness to use a BTE processor for the duration of the study ARH Arm Inclusion Criteria * Residual low frequency hearing sensitivity (pure tone average of < 70 dB HL for 125, 250, and 500 Hz) and a severe-to-profound high-frequency sensorineural hearing loss (pure tone average of ≥ 70 dB HL for 1,000, 2,000, 3000, 4,000, and 8,000 Hz) in the implanted ear * Willingness to use an in-canal acoustic earhook for the duration of the study EO Arm Inclusion Criteria • Severe-to-profound sensorineural hearing loss in the low (pure tone average of ≥ 70 dB HL for 125, 250, and 500 Hz) and high frequencies (pure tone average ≥ 70 dB HL for 1,000, 2,000, 3000, 4,000, and 8,000 Hz) in the implanted ear Exclusion Criteria: * Unrealistic expectations regarding potential benefits, risks and limitations of the investigational device as determined by the investigator * Unwillingness or physical / cognitive inability of subject to comply with all investigational requirements as determined by the investigator
| Milestone | Electric Only (EO) Cohort | Aidable Residual Hearing (ARH) Cohort |
|---|---|---|
| Started | 10 | 12 |
| Completed | 10 | 12 |
| Not completed | 0 | 0 |
Time frame: AEs were recorded and tracked between completion of the informed consent form (signed and dated) and two weeks after the participant's last study visit. The total period is up to 5 weeks.
| Arm | Deaths | Serious adverse events | Other adverse events |
|---|---|---|---|
| EO Cohort | 0/10 (0.0%) | 0/10 (0.0%) | 2/10 (20.0%) |
| ARH Cohort | 0/12 (0.0%) | 0/12 (0.0%) | 1/12 (8.3%) |
| Term | Organ system | EO Cohort | ARH Cohort |
|---|---|---|---|
| Patient discomfort | Product Issues | 0/10 | 1/12 |
| Upper Respiratory | Infections and infestations | 1/10 | 0/12 |
| Sprained ankle | Musculoskeletal and connective tissue disorders | 1/10 | 0/12 |
Limitations and caveats: Due to clinic closures related to the COVID-19 pandemic, total enrollment in the EO Cohort was 10 subjects instead of 12.
| Facility | City | Country | Status |
|---|---|---|---|
| University of Colorado Anschutz Medical Campus | Aurora, Colorado | United States | |
| Tampa Bay Hearing and Balance Center | Tampa, Florida | United States | |
| Midwest Ear Institute/St. Luke's Health System | Kansas City, Missouri | United States | |
| Washington University Medical Center | St Louis, Missouri | United States | |
| Vanderbilt University Medical Center | Nashville, Tennessee | United States | |
| Austin Ear, Nose & Throat Clinic | Austin, Texas | United States |