Safety and Feasibility of Liver Retraction With the Levita Magnetic Surgical System: Extended Magnetic Grasper Device

NCT06432088 — Not applicable, Completed, 30 participants, Levita Magnetics

ClinicalTrials Daily  ·  Areas  ·  Conditions  ·  Drugs  ·  Sponsors  ·  

Prospective, Multi-center, Single-arm, Open Label Study Designed to Assess the Safety and Feasibility of Liver Retraction With the Levita Magnetic Surgical System: Extended Magnetic Grasper Device

RegistryNCT06432088
StatusCompleted
PhaseNot applicable
Study typeInterventional; na, single group, masking: none, other
Enrollment30 (Actual)
SponsorLevita Magnetics (Industry)
ConditionsBariatric Surgery Candidate, Hiatal Hernia
InterventionsExtended Magnetic Grasper Device (Device)
Datesstart 2024-05-13, primary completion 2024-09-28, completion 2024-09-30, first posted 2024-05-29, last updated 2025-05-29
Locations1 site in Chile

Related Papers Related Patents Related Grants Related Experts
All trials for Bariatric Surgery Candidate All trials for Hiatal Hernia All trials of Extended Magnetic Grasper Device

-Summary

The purpose of this study is to evaluate the safety and effectiveness of the Extended Magnetic Grasper Device in patients undergoing bariatric and/or hiatal hernia procedures, as a liver retractor grasping the liver and/or the tissue surrounding the crus of the diaphragm.

Expand all sections  ·  Collapse all

+Detailed description

The purpose of this study is to evaluate the safety and effectiveness of the Levita Magnetic Surgical System Version 2 in patients undergoing bariatric and/or hiatal hernia procedures.

+Arms and interventions

ArmTypeInterventionsDescription
Extended Magnetic Grasper DeviceExperimentalDevice: Extended Magnetic Grasper DeviceEvaluation of the safety and effectiveness of the Extended Magnetic Grasper Device in patients undergoing bariatric and/or hiatal hernia procedures, as a liver retractor grasping the liver and/or the tissue surrounding the crus of the diaphragm.
  • Extended Magnetic Grasper Device (Device): The purpose of this study is to evaluate the safety and effectiveness of the Extended Magnetic Grasper Device in patients undergoing bariatric and/or hiatal hernia procedures, as a liver retractor grasping the liver and/or the tissue surrounding the crus of the diaphragm.

+Outcome measures

Primary outcomes

  • Number of Device Related Adverse Events [30 days]: All adverse events will be captured and reported. Adverse events will be summarized by relatedness to the device and/or procedure, seriousness and level of severity.
  • Mobilization Requiring Additional Retractors [1 day]: Ability to adequately mobilize the liver to achieve an effective exposure of the target tissue. Adequate mobilization is not achieved if it is necessary to use another liver retractor during the procedure.

+Eligibility

Sex: All, ages: 18 Years to any

Inclusion Criteria:

* At least 18 years of age.
* Scheduled to undergo elective bariatric and/or hiatal hernia procedures.
* Willing and able to provide a written Informed Consent Form (ICF) to participate in the study prior to any study required procedures.

Exclusion Criteria:

* Individuals with pacemakers, defibrillators, or other electromedical implants.
* Individuals with ferromagnetic implants.
* American Society of Anesthesiologists (ASA) score of III or IV.
* Significant comorbidities: cardiovascular, neuromuscular, chronic obstructive pulmonary disease, and urological disease (renal failure).
* Clinical history of impaired coagulation confirmed by abnormal blood tests.
* Individuals has signs of hepatic abnormality (e.g.: cirrhosis, liver failure, increase in liver enzymes, etc.).
* Anatomical abnormality or disease of intended target tissue noted after initiation of index procedure that would prevent device use.
* Pregnant or wishes to become pregnant during the length of study participation.
* Individual is not likely to comply with the follow-up evaluation schedule.
* Participating in a clinical trial of another investigational drug or device.
* Prisoner or under incarceration.

+Posted results

Participant flow

MilestoneExtended Magnetic Grasper Device
Started30
Completed30
Not completed0

Outcome results

  • Number of Device Related Adverse Events (Primary), Number, Device related adverse events: Extended Magnetic Grasper Device 0
  • Mobilization Requiring Additional Retractors (Primary), Number, Procedures w/ additional retractors: Extended Magnetic Grasper Device 0

Adverse events

Time frame: Adverse events (AE) and serious adverse events (SAEs) were recorded from index procedure through study exit (30 day follow up).

ArmDeathsSerious adverse eventsOther adverse events
Extended Magnetic Grasper Device0/30 (0.0%)0/30 (0.0%)14/30 (46.7%)

Other adverse events by term

TermOrgan systemExtended Magnetic Grasper Device
Ecchymosis in Surgical IncisionSkin and subcutaneous tissue disorders5/30
Liver Capsule Minor AbrasionSurgical and medical procedures5/30
PetechiaSkin and subcutaneous tissue disorders7/30
Sinus TachycardiaCardiac disorders1/30
Transverse Colon InjuryGastrointestinal disorders1/30
Atrioventricular BlockCardiac disorders1/30

+Locations (1)

FacilityCityCountryStatus
Hospital Santiago Oriente Dr. Luis Tisné BrousseSantiago, Santiago MetropolitanChile

+Study documents