In Vivo Testing for SpO2 Accuracy on Healthy Adults Under Laboratory Conditions
| Registry | NCT05873335 |
|---|---|
| Status | Completed |
| Phase | Not applicable |
| Study type | Interventional; na, single group, masking: none, other |
| Enrollment | 12 (Actual) |
| Sponsor | Gabi SmartCare (Industry) |
| Conditions | Hypoxia |
| Interventions | Mixtures of nitrogen, air, and carbon dioxide (Combination product) |
| Dates | start 2023-02-28, primary completion 2023-03-01, completion 2023-03-01, first posted 2023-05-24, last updated 2024-12-12 |
| Locations | 1 site in United States |
Related Papers Related Patents Related Grants Related Experts
All trials for Hypoxia All trials of Mixtures of nitrogen, air, and carbon dioxide
The goal of this clinical trial is to compare values of SpO2 provided by the DUT and the SaO2 used as a reference in healthy adults. The main question it aims to answer is: what is the accuracy of the subject device compared to the 510(k)-cleared reference device
A radial arterial cannula was placed in the wrist of each subject for arterial blood sampling and a subject device was placed on each arm of the patients. Then, hypoxia was induced to different stable levels of oxyhemoglobin saturation between 70-100% by having subjects breathe mixtures of nitrogen, air, and carbon dioxide controlled by the study physician. Values of SpO2 and SaO2 were then compared to provide a root mean square error and conclude on the subject device accuracy.
| Arm | Type | Interventions | Description |
|---|---|---|---|
| Healthy Adults | Experimental | Combination Product: Mixtures of nitrogen, air, and carbon dioxide | Healthy adults |
Sex: All, ages: 21 Years to 50 Years, accepts healthy volunteers
Inclusion Criteria: * Healthy adult * Normal hemoglobin levels (Hemoglobin > 10 g/dL) Exclusion Criteria: * Systemic illness * Obesity
| Milestone | Healthy Adults |
|---|---|
| Started | 12 |
| Completed | 12 |
| Not completed | 0 |
Time frame: 2 hours
| Arm | Deaths | Serious adverse events | Other adverse events |
|---|---|---|---|
| Healthy Adults | 0/12 (0.0%) | 0/12 (0.0%) | 0/12 (0.0%) |
| Facility | City | Country | Status |
|---|---|---|---|
| Hypoxia Research Laboratory | San Francisco, California | United States |