Clinical Evaluation of a Small Aperture Extended Depth of Focus Intraocular Lens

NCT03633695 — Not applicable, Completed, 453 participants, AcuFocus, Inc.

ClinicalTrials Daily  ·  Areas  ·  Conditions  ·  Drugs  ·  Sponsors  ·  
RegistryNCT03633695
StatusCompleted
PhaseNot applicable
Study typeInterventional; non randomized, parallel, masking: single, treatment
Enrollment453 (Actual)
SponsorAcuFocus, Inc. (Industry)
ConditionsCataract, Presbyopia
InterventionsIC-8 IOL Group, Control Group (Device)
Datesstart 2018-12-04, primary completion 2020-10-15, completion 2020-10-15, first posted 2018-08-16, last updated 2022-10-24
Locations21 sites in United States

Related Papers Related Patents Related Grants Related Experts
All trials for Cataract All trials for Presbyopia All trials of IC-8 IOL Group All trials of Control Group

-Summary

The purpose of this study is to determine the safety and effectiveness of the IC-8 IOL implanted in one eye and a monofocal or monofocal toric IOL implanted in the fellow eye in accordance with the indication.

Expand all sections  ·  Collapse all

+Detailed description

Eligible patients will complete up to 12 study visits over a 12-month period. All subjects will complete a preoperative examination of both eyes to assess study eligibility: operative visit (each eye) and up to 9 postoperative visits (Day 1 - each eye, Week 1 - each eye, Month 1 - each eye or combined, Month 3 - both eyes, Month 6 - both eyes, and Month 12 - both eyes). The first eye must meet specific qualification criteria in order to proceed with IOL implantation in the second eye. The second eye should be implanted within 45 days of the first eye.

+Arms and interventions

ArmTypeInterventionsDescription
IC-8 IOL GroupExperimentalDevice: IC-8 IOL GroupA monofocal IOL (TECNIS ZCB00, ZCT150 or ZCT225 or AcrySof IQ SA60WF, SA6AT3 or SA6AT4) implanted in the first eye. The AcuFocus IC-8 IOL implanted in the second eye.
Control GroupActive comparatorDevice: Control GroupA monofocal IOL (TECNIS ZCB00, ZCT150 or ZCT225 or AcrySof IQ SA60WF, SA6AT3 or SA6AT4) bilaterally implanted.
  • IC-8 IOL Group (Device), also known as Intraocular Lens: A monofocal IOL (TECNIS ZCB00, ZCT150 or ZCT225 or AcrySof IQ SA60WF, SA6AT3 or SA6AT4) implanted in the first eye for use over the lifetime of the patient. The IC-8 IOL implanted in the second eye for use over the lifetime of the patient.
  • Control Group (Device), also known as Intraocular lens: A monofocal IOL (TECNIS ZCB00, ZCT150 or ZCT225 or AcrySof IQ SA60WF, SA6AT3 or SA6AT4) bilaterally implanted for lifetime use of the patient.

+Outcome measures

Primary outcomes

  • Mean Binocular Photopic Uncorrected Intermediate (66 cm) Visual Acuity (UCIVA) [6 Months (160-210 days post second eye operative visit)]: Visual acuity (VA) was tested binocularly (both eyes) at a distance of 66 cm, without manifest refraction, using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart.
  • Mean Binocular Photopic Uncorrected Near (40 cm) Visual Acuity (UCNVA) [6 Months (160-210 days post second eye operative visit)]: Visual acuity (VA) was tested binocularly (both eyes) at a distance of 40 cm, without manifest refraction, using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart.
  • Mean Binocular Photopic Uncorrected Distance (4 m) Visual Acuity (UCDVA) [6 Months (160-210 days post second eye operative visit)]: Visual acuity (VA) was tested binocularly (both eyes) at a distance of 4 m, with +0.25 D infinity adjustment lens in front of the eyes, using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart.
  • Mean Monocular Photopic Distance-Corrected Intermediate (66 cm) Visual Acuity (DCIVA) in IC-8™ IOL Eyes [6 Months (160-210 days post second eye operative visit)]: Visual acuity (VA) was tested monocularly (each eye) at a distance of 66 cm, with the distance manifest refraction (with infinity adjustment) in place for that eye, using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart.
  • Mean Monocular Photopic Depth of Focus (DOF) in IC-8™ IOL Eyes [3 Months (60-110 days post second eye operative visit)]: Depth of focus was performed monocularly in the IC-8™ IOL Group at the far to near range of vision (+2.00 to -5.00 D), using 100% contrast ETDRS charts in a computerized testing system calibrated for a 4 m test distance, with manifest refraction (no infinity adjustment) in place for the eye(s) being tested. The defocus power was progressively introduced in 0.50 D increments from +2.00 D to +0.50 D and from -0.50 D to -5.00 D, then in 0.25 D increments from +0.50 D to -0.50 D, while visual acuity was measured at each successive defocus step. The depth of focus was estimated as the defocus range between zero defocus and the first point on the negative lens induced mean defocus curve that crosses the 0.2 logMAR using a linear interpolation.
  • Mean Monocular Best-corrected Distance (4 m) Visual Acuity (BCDVA) [12 Months (300-420 days post second eye operative visit)]: Visual acuity (VA) was tested monocularly (each eye) at a distance of 4 m, in photopic conditions, with manifest refraction in place (no infinity adjustment) using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart.
  • Proportion of IC-8™ IOL Eyes Achieving Best-corrected Distance Visual Acuity (BCDVA) 0.3 logMAR or Better Compared to the Safety and Performance Endpoints (SPE) Rates [12 Months (300-420 days post second eye operative visit)]: Visual acuity (VA) was tested monocularly (each eye) at a distance of 4 m, in photopic conditions, with manifest refraction in place (no infinity adjustment) using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart.
  • Proportion of Best-Case IC-8™ IOL Eyes Achieving Best-corrected Distance Visual Acuity (BCDVA) 0.3 logMAR or Better Compared to the Safety and Performance Endpoints (SPE) Rates [12 Months (300-420 days post second eye operative visit)]: Visual acuity (VA) was tested monocularly (each eye) at a distance of 4 m, in photopic conditions, with manifest refraction in place (no infinity adjustment) using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart. Monocular BCDVA for the IC-8™ IOL eyes of subjects in the Best-Case population was compared to the safety and performance endpoints (SPE) rate for posterior chamber IOLs.
  • Rates of IC-8™ IOL Eyes With Cumulative Ocular Serious Adverse Events [Through 12 Months (300-420 days post second eye operative visit)]: The rates of cumulative ocular serious adverse events (SAEs) for IC-8™ IOL eyes were calculated from the time of second eye (IC-8™ IOL eye) implantation through Month 12.
  • Rates of IC-8™ IOL Eyes With Cumulative Postoperative Ocular Adverse Events (Secondary Surgical Interventions) [Through 12 Months (300-420 days post second eye operative visit)]: The rates of cumulative ocular adverse events for IC-8™ IOL eyes were calculated from the time of second eye (IC-8™ IOL eye) implantation through Month 12.
  • Rates of IC-8™ IOL Eyes With Persistent Ocular Serious Adverse Events [Through 12 Months (300-420 days post second eye operative visit)]: The rates of persistent (defined as being unresolved at final scheduled visit) ocular adverse events (AEs) for IC-8™ IOL eyes were calculated from the time of second eye (IC-8™ IOL eye) implantation through Month 12.
  • Rates of IC-8™ IOL Eyes With IC-8™ IOL Removals [Through 12 Months (300-420 days post second eye operative visit)]: The rates of cumulative ocular adverse events for IC-8™ IOL eyes were calculated from the time of second eye (IC-8™ IOL eye) implantation through Month 12.
  • Rates of Eyes With Postoperative Ocular Adverse Events (Other Than Secondary Surgical Intervention), Related to Device (Serious and Non-Serious Combined) [Through 12 Months (300-420 days post second eye operative visit)]: The rates of cumulative ocular adverse events (AEs) for IC-8™ IOL eyes were calculated from the time of second eye (IC-8™ IOL eye) implantation through Month 12.

Secondary outcomes

  • Tolerance to Preoperative Corneal Astigmatism in IC-8™ IOL Eyes With BCDVA 20/25 at 3 Months [3 Months (60-110 days post second eye operative visit)]: Uncorrected-distance visual acuity (UCDVA) in IC-8™ IOL eyes achieving best-corrected distance visual acuity (BCDVA) of 20/25 or better was compared between eyes with preoperative corneal astigmatism < 1.0 D to eyes with 1.0 D to 1.5 D of preoperative corneal astigmatism.
  • Mean Monocular Photopic Contrast Sensitivity Without Glare [6 Months (160-210 days post second eye operative visit)]: Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested monocularly (each eye) in photopic (well-lit) conditions, without a glare source, using sine waving gratings in a computerized testing system. The photopic spatial frequencies tested were 3, 6, 12, 18 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better contrast sensitivity results are reflected by a higher numeric value.
  • Mean Monocular Photopic Contrast Sensitivity With Glare [6 Months (160-210 days post second eye operative visit)]: Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects, and was tested monocularly (each eye) in photopic (well-lit) conditions, with a glare source, using sine waving gratings in a computerized testing system. The photopic spatial frequencies tested were 3, 6, 12, 18 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.
  • Mean Monocular Mesopic Contrast Sensitivity Without Glare [6 Months (160-210 days post second eye operative visit)]: Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested monocularly (each eye) in mesopic (low-light) conditions, without a glare source, using sine waving gratings in a computerized testing system. The mesopic spatial frequencies tested were 1.5, 3, 6, 12 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.
  • Mean Monocular Mesopic Contrast Sensitivity With Glare [6 Months (160-210 days post second eye operative visit)]: Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested monocularly (each eye) in mesopic (low-light) conditions, with a glare source, using sine waving gratings in a computerized testing system. The mesopic spatial frequencies tested were 1.5, 3, 6, 12 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.
  • Mean Binocular Photopic Contrast Sensitivity Without Glare [6 Months (160-210 days post second eye operative visit)]: Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested binocularly (both eyes) in photopic (well-lit) conditions, without a glare source, using sine waving gratings in a computerized testing system. The photopic spatial frequencies tested were 3, 6, 12, 18 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.
  • Mean Binocular Photopic Contrast Sensitivity With Glare [6 Months (160-210 days post second eye operative visit)]: Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested binocularly (both eyes) in photopic (well-lit) conditions, with a glare source, using sine waving gratings in a computerized testing system. The photopic spatial frequencies tested were 3, 6, 12, 18 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.
  • Mean Binocular Mesopic Contrast Sensitivity Without Glare [6 Months (160-210 days post second eye operative visit)]: Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested binocularly (both eyes) in mesopic (low-light) conditions, without a glare source, using sine waving gratings in a computerized testing system. The mesopic spatial frequencies tested were 1.5, 3, 6, 12 cycles per degree (cpd). When a subject was unable to see the highest contrast, the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.
  • Mean Binocular Mesopic Contrast Sensitivity With Glare [6 Months (160-210 days post second eye operative visit)]: Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested binocularly (both eyes) in mesopic (low-light) conditions, with a glare source, using sine waving gratings in a computerized testing system. The mesopic spatial frequencies tested were 1.5, 3, 6, 12 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.

+Eligibility

Sex: All, ages: 22 Years to any, accepts healthy volunteers

Inclusion Criteria:

1. Minimum 22 years of age;
2. Able to comprehend and have signed a statement of informed consent;
3. Availability, willingness, ability and sufficient cognitive awareness to comply with examination procedures and study visits;
4. Planned crystalline lens removal by phacoemulsification, with or without femtosecond laser-assisted extraction, and posterior chamber IOL implantation in both eyes;
5. Cataractous lens changes as demonstrated by best-corrected visual acuity (BCDVA) of 20/40 or worse either with or without a glare source present;
6. Potential for postoperative BCDVA of 20/25 or better in each eye
7. Clear intraocular media, other than cataract.

Exclusion Criteria:

1. Requiring an IC-8 intraocular lens outside the available spherical power range
2. Pharmacologically dilated pupil size less than 6 mm in either eye;
3. Inability to achieve stable keratometric readings for contact lens wearers
4. Irregular astigmatism in either eye;
5. Preoperative corneal astigmatism > 1.50 diopters in either eye
6. Active or recurrent anterior segment pathology
7. Presence of ocular abnormalities other than cataract as specified in the protocol
8. Diagnosis of dry eye in which patients are unable to maintain eye comfort or adequate vision even with dry eye medication;
9. Congenital cataracts;
10. Previous corneal or intraocular surgery
11. History of ocular trauma or ocular conditions expected to require retinal laser treatment or other surgical intervention;
12. Systemic conditions as specified in the protocol;
13. Patient is pregnant, plans to become pregnant, or is lactating
14. Concurrent participation or participation in any clinical trial up to 30 days prior to preoperative visit.

+Posted results

Participant flow

MilestoneIC-8™ IOL GroupControl Group
Started343110
Astigmatism group 1 (ic-8™ iol group)2730
Astigmatism group 2 (ic-8™ iol group)700
Astigmatism group 1 (control group)091
Astigmatism group 2 (control group)019
Completed331101
Not completed129

Outcome results

  • Mean Binocular Photopic Uncorrected Intermediate (66 cm) Visual Acuity (UCIVA) (Primary), Mean, logMAR: IC-8™ IOL Group 0.051 +/- 0.1629; Control Group 0.228 +/- 0.1646 (p <.0001)
  • Mean Binocular Photopic Uncorrected Near (40 cm) Visual Acuity (UCNVA) (Primary), Mean, logMAR: IC-8™ IOL Group 0.186 +/- 0.1425; Control Group 0.377 +/- 0.1576 (p <.0001)
  • Mean Binocular Photopic Uncorrected Distance (4 m) Visual Acuity (UCDVA) (Primary), Mean, logMAR: IC-8™ IOL Group -0.010 +/- 0.1063; Control Group 0.002 +/- 0.0992 (p <.0001)
  • Mean Monocular Photopic Distance-Corrected Intermediate (66 cm) Visual Acuity (DCIVA) in IC-8™ IOL Eyes (Primary), Mean, logMAR: IC-8™ IOL Eyes (IC-8™ IOL Group) 0.144 +/- 0.1709; Fellow Eyes (IC-8™ IOL Group) 0.325 +/- 0.1687 (p <.0001)
  • Mean Monocular Photopic Depth of Focus (DOF) in IC-8™ IOL Eyes (Primary), Mean, Diopter: IC-8™ IOL Eyes (IC-8™ IOL Group) -1.99 +/- NA; Fellow Eyes (IC-8™ IOL Group) -1.08 +/- NA
  • Mean Monocular Best-corrected Distance (4 m) Visual Acuity (BCDVA) (Primary), Mean, logMAR: IC-8™ IOL Eyes (IC-8™ IOL Group) 0.009 +/- 0.1131; Fellow Eyes (IC-8™ IOL Group) -0.059 +/- 0.0928 (p <.0001)
  • Proportion of IC-8™ IOL Eyes Achieving Best-corrected Distance Visual Acuity (BCDVA) 0.3 logMAR or Better Compared to the Safety and Performance Endpoints (SPE) Rates (Primary), Count of units, Eyes: IC-8™ IOL Eyes (IC-8™ IOL Group) 326; Fellow Eyes (IC-8™ IOL Group) 330
  • Proportion of Best-Case IC-8™ IOL Eyes Achieving Best-corrected Distance Visual Acuity (BCDVA) 0.3 logMAR or Better Compared to the Safety and Performance Endpoints (SPE) Rates (Primary), Count of units, Eyes: IC-8™ IOL Eyes (IC-8™ IOL Group) 315; Fellow Eyes (IC-8™ IOL Group) 318
  • Rates of IC-8™ IOL Eyes With Cumulative Ocular Serious Adverse Events (Primary), Count of units, Eyes: IC-8™ IOL Eyes (IC-8™ IOL Group) 5; Fellow Eyes (IC-8™ IOL Group) 4
  • Rates of IC-8™ IOL Eyes With Cumulative Postoperative Ocular Adverse Events (Secondary Surgical Interventions) (Primary), Count of units, Eyes: IC-8 IOL Eyes (IC-8™ IOL Group) 1; Fellow Eyes (IC-8™ IOL Group) 1; Second Eyes (Control Group) 0; First Eyes (Control Group) 1
  • Rates of IC-8™ IOL Eyes With Persistent Ocular Serious Adverse Events (Primary), Count of units, Eyes: IC-8™ IOL Eyes (IC-8™ IOL Group) 0; Fellow Eyes (IC-8™ IOL Group) 0
  • Rates of IC-8™ IOL Eyes With IC-8™ IOL Removals (Primary), Count of units, Eyes: IC-8™ IOL Eyes (IC-8™ IOL Group) 0; Fellow Eyes (IC-8™ IOL Group) 0; Second Eye (Control Group) 1; First Eye (Control Group) 1
  • Rates of Eyes With Postoperative Ocular Adverse Events (Other Than Secondary Surgical Intervention), Related to Device (Serious and Non-Serious Combined) (Primary), Count of units, Eyes: IC-8 IOL Eyes (IC-8™ IOL Group 1; Fellow Eyes (IC-8™ IOL Group 0; Second Eyes (Control Group) 0; First Eyes (Control Group) 0
  • Tolerance to Preoperative Corneal Astigmatism in IC-8™ IOL Eyes With BCDVA 20/25 at 3 Months (Secondary), Mean, logMAR: IC-8™ IOL Eyes With Astigmatism <1.0 D (IC-8™ IOL Group) 0.085 +/- 0.1269; IC-8™ IOL Eyes With Astigmatism 1.0 D - 1.5 D (IC-8™ IOL Group) 0.108 +/- 0.1208 (p <.0001)
  • Mean Monocular Photopic Contrast Sensitivity Without Glare (Secondary), Mean, log units on a scale: IC-8™ IOL Eyes (IC-8™ IOL Group) 1.935 +/- 0.3768; Fellow Eyes (IC-8™ IOL Group) 2.202 +/- 0.2542
  • Mean Monocular Photopic Contrast Sensitivity With Glare (Secondary), Mean, log units on a scale: IC-8™ IOL Eyes (IC-8™ IOL Group) 1.620 +/- 0.4219; Fellow Eyes (IC-8™ IOL Group) 1.822 +/- 0.4247
  • Mean Monocular Mesopic Contrast Sensitivity Without Glare (Secondary), Mean, log units on a scale: IC-8™ IOL Eyes (IC-8™ IOL Group) 1.385 +/- 0.3596; Fellow Eyes (IC-8™ IOL Group) 1.850 +/- 0.3849
  • Mean Monocular Mesopic Contrast Sensitivity With Glare (Secondary), Mean, log units on a scale: IC-8™ IOL Eyes (IC-8™ IOL Group) 1.044 +/- 0.4063; Fellow Eyes (IC-8™ IOL Group) 1.338 +/- 0.4212
  • Mean Binocular Photopic Contrast Sensitivity Without Glare (Secondary), Mean, log units on a scale: IC-8™ IOL Group 2.233 +/- 0.2536; Control Group 2.287 +/- 0.1896
  • Mean Binocular Photopic Contrast Sensitivity With Glare (Secondary), Mean, log units on a scale: IC-8™ IOL Group 1.866 +/- 0.3893; Control Group 1.983 +/- 0.3217
  • 2 more outcome measures not shown.

Adverse events

Time frame: Second eye implantation through study completion; approximately 1-year (12 months).

ArmDeathsSerious adverse eventsOther adverse events
Preoperative0/453 (0.0%)0/453 (0.0%)0/453 (0.0%)
IC-8™ IOL Group - 2nd Eye0/343 (0.0%)6/343 (1.7%)264/343 (77.0%)
IC-8™ IOL Group - 1st Eye0/343 (0.0%)2/343 (0.6%)148/343 (43.1%)
IC-8™ IOL Group - Systemic0/343 (0.0%)22/343 (6.4%)114/343 (33.2%)
Control Group - 2nd Eye0/110 (0.0%)1/110 (0.9%)56/110 (50.9%)
Control Group - 1st Eye0/110 (0.0%)1/110 (0.9%)55/110 (50.0%)
Control Group - Systemic0/110 (0.0%)5/110 (4.5%)50/110 (45.5%)

Serious adverse events by term

TermOrgan systemPreoperativeIC-8™ IOL Group - 2nd EyeIC-8™ IOL Group - 1st EyeIC-8™ IOL Group - SystemicControl Group - 2nd EyeControl Group - 1st EyeControl Group - Systemic
AbscessInfections and infestations0/4530/3430/3431/3430/1100/1100/110
AneursysmBlood and lymphatic system disorders0/4530/3430/3432/3430/1100/1100/110
ArrhythmiaVascular disorders0/4530/3430/3432/3430/1100/1100/110
ArthritisMusculoskeletal and connective tissue disorders0/4530/3430/3431/3430/1100/1100/110
Electrolyte imbalanceMetabolism and nutrition disorders0/4530/3430/3431/3430/1100/1100/110
Heart failure/Congestive heart failureVascular disorders0/4530/3430/3431/3430/1100/1100/110
Hemorrhage/BleedingBlood and lymphatic system disorders0/4530/3430/3430/3430/1100/1101/110
Ischemic heart diseaseVascular disorders0/4530/3430/3434/3430/1100/1101/110
PancreatitisGastrointestinal disorders0/4530/3430/3430/3430/1100/1101/110
PneumoniaRespiratory, thoracic and mediastinal disorders0/4530/3430/3431/3430/1100/1100/110
Pulmonary embolismVascular disorders0/4530/3430/3431/3430/1100/1101/110
Respiratory failureRespiratory, thoracic and mediastinal disorders0/4530/3430/3431/3430/1100/1101/110
SepsisInfections and infestations0/4530/3430/3431/3430/1100/1100/110
Solid tumourImmune system disorders0/4530/3430/3431/3430/1100/1100/110
Spinal column injuryMusculoskeletal and connective tissue disorders0/4530/3430/3431/3430/1100/1100/110
Suicidal ideationPsychiatric disorders0/4530/3430/3431/3430/1100/1100/110
Transient ischemic attackVascular disorders0/4530/3430/3431/3430/1100/1100/110
Urinary tract infectionInfections and infestations0/4530/3430/3431/3430/1100/1100/110
VomitingGeneral disorders0/4530/3430/3431/3430/1100/1100/110
Cystoid Macular EdemaEye disorders0/4533/3432/3430/3431/1100/1100/110
EndophthalmitisEye disorders0/4531/3430/3430/3430/1100/1100/110
Intraocular Pressure IncreasedEye disorders0/4531/3430/3430/3430/1100/1100/110
Iris InjuryEye disorders0/4530/3430/3430/3430/1101/1100/110
Retinal Vein OcclusionEye disorders0/4531/3430/3430/3430/1100/1100/110

Other adverse events by term

TermOrgan systemPreoperativeIC-8™ IOL Group - 2nd EyeIC-8™ IOL Group - 1st EyeIC-8™ IOL Group - SystemicControl Group - 2nd EyeControl Group - 1st EyeControl Group - Systemic
Abdominal painGeneral disorders0/4530/3430/3431/3430/1100/1100/110
AbscessInfections and infestations0/4530/3430/3432/3430/1100/1101/110
Acute liver failureHepatobiliary disorders0/4530/3430/3430/3430/1100/1101/110
Altered Visual Depth PerceptionEye disorders0/4531/3431/3430/3430/1100/1100/110
AnemiaBlood and lymphatic system disorders0/4530/3430/3430/3430/1100/1101/110
AngioedemaMusculoskeletal and connective tissue disorders0/4530/3430/3432/3430/1100/1100/110
Anterior Capsular PhimosisEye disorders0/4531/3430/3430/3430/1100/1100/110
AnxietyGeneral disorders0/4530/3430/3432/3430/1100/1101/110
ArrhythmiaVascular disorders0/4530/3430/3431/3430/1100/1100/110
Arteriosclerosis/AtherosclerosisBlood and lymphatic system disorders0/4530/3430/3431/3430/1100/1100/110
ArthralgiaMusculoskeletal and connective tissue disorders0/4530/3430/3432/3430/1100/1100/110
Arthralgia/MyalgiaMusculoskeletal and connective tissue disorders0/4530/3430/3431/3430/1100/1100/110
ArthritisMusculoskeletal and connective tissue disorders0/4530/3430/3432/3430/1100/1100/110
AsthmaRespiratory, thoracic and mediastinal disorders0/4530/3430/3431/3430/1100/1100/110
Atrial fibrillationVascular disorders0/4530/3430/3432/3430/1100/1101/110
BradycardiaSurgical and medical procedures0/4530/3430/3432/3430/1100/1100/110
Breast cancerImmune system disorders0/4530/3430/3431/3430/1100/1100/110
BronchitisRespiratory, thoracic and mediastinal disorders0/4530/3430/3431/3430/1100/1102/110
Bronchospasm/AsthmaRespiratory, thoracic and mediastinal disorders0/4530/3430/3431/3430/1100/1100/110
Brow, Eyelid PtosisEye disorders0/4532/3432/3430/3430/1100/1100/110
CancerImmune system disorders0/4530/3430/3435/3430/1100/1100/110
Cataract Fragments in Eye Following Cataract SurgeryEye disorders0/4532/3431/3430/3430/1100/1100/110
Cataract Operation Complication, Lens DislocationEye disorders0/4531/3431/3430/3430/1101/1100/110
Cataract SubcapsularEye disorders0/4531/3430/3430/3430/1100/1100/110
Catheterization arterial normalVascular disorders0/4530/3430/3430/3430/1100/1101/110
Ciliary Zonular Dehiscence, Vitreous ProlapseEye disorders0/4530/3430/3430/3431/1100/1100/110
Cold soresImmune system disorders0/4530/3430/3431/3430/1100/1100/110
Common cold syndromeGeneral disorders0/4530/3430/3431/3430/1100/1100/110
Conjunctival EdemaEye disorders0/4530/3431/3430/3430/1100/1100/110
Conjunctival HemorrhageEye disorders0/4530/3433/3430/3430/1100/1100/110
Conjunctival abrasionEye disorders0/4530/3430/3430/3431/1100/1100/110
ConjunctivitisEye disorders0/4537/3434/3430/3432/1101/1100/110
Convulsion/SeizureNervous system disorders0/4530/3430/3430/3430/1100/1101/110
Corneal AbrasionEye disorders0/4531/3430/3430/3432/1102/1100/110
Corneal Deposits, Cornea VerticillataEye disorders0/4531/3431/3430/3430/1100/1100/110
Corneal DystrophyEye disorders0/4533/3433/3430/3430/1100/1100/110
Corneal Epithelial DefectEye disorders0/4531/3430/3430/3430/1100/1100/110
Corneal Epithelial MicrocystsEye disorders0/4531/3431/3430/3430/1100/1100/110
Corneal stromal edemaEye disorders0/4530/3430/3430/3431/1100/1100/110
Cramp(s)/Muscle spasm(s)Musculoskeletal and connective tissue disorders0/4530/3430/3433/3430/1100/1100/110
Cystoid Macular EdemaEye disorders0/4532/3433/3430/3430/1101/1100/110
Dacryostenosis acquiredEye disorders0/4531/3431/3430/3430/1100/1100/110
Damage to ligament(s)Musculoskeletal and connective tissue disorders0/4530/3430/3431/3430/1100/1100/110
DehydrationMetabolism and nutrition disorders0/4530/3430/3431/3430/1100/1100/110
DepressionGeneral disorders0/4530/3430/3432/3430/1100/1100/110
Detached Descemet's membraneEye disorders0/4532/3430/3430/3430/1100/1100/110
Detachment of retinal pigment epithelium, Retinal depigmentationEye disorders0/4530/3430/3430/3431/1101/1100/110
Device DamageEye disorders0/4532/3430/3430/3430/1100/1100/110
DiplopiaEye disorders0/4531/3432/3430/3430/1100/1100/110
DizzinessGeneral disorders0/4530/3430/3431/3430/1100/1101/110
Dry EyeEye disorders0/45332/34326/3430/3439/1109/1100/110
DysphotopsiaEye disorders0/4535/3433/3430/3431/1102/1100/110
EczemaSkin and subcutaneous tissue disorders0/4530/3430/3431/3430/1100/1100/110
Electrolyte ImbalanceMetabolism and nutrition disorders0/4530/3430/3431/3430/1100/1100/110
EncephalopathyInfections and infestations0/4530/3430/3430/3430/1100/1101/110
Eye IrritationEye disorders0/4531/3431/3430/3431/1100/1100/110
Eye PainEye disorders0/4535/3431/3430/3431/1101/1100/110
Eye PruritusEye disorders0/4532/3432/3430/3430/1100/1100/110
Eyelid infectionEye disorders0/4536/3432/3430/3431/1100/1100/110
Eyelid tumorEye disorders0/4531/3430/3430/3430/1100/1100/110

77 more terms not shown.

+Locations (21)

FacilityCityCountryStatus
Fishkind, Bakewell, Maltzman, Hunter & Associates Eye Care & Surgery CenterTucson, ArizonaUnited States
Empire Eye & Laser CenterBakersfield, CaliforniaUnited States
Harvard Eye AssociatesLaguna Hills, CaliforniaUnited States
Altos Eye PhysiciansLos Altos, CaliforniaUnited States
Advanced Vision CareLos Angeles, CaliforniaUnited States
Eye Center of Northern ColoradoFort Collins, ColoradoUnited States
Eye Center of North FloridaPanama City, FloridaUnited States
Chu Vision InstituteBloomington, MinnesotaUnited States
Pepose Vision InstituteSt Louis, MissouriUnited States
Kugler VisionOmaha, NebraskaUnited States
Alterman, Modi & WolterPoughkeepsie, New YorkUnited States
Physicians ProtocolGreensboro, North CarolinaUnited States
Cleveland Eye ClinicBrecksville, OhioUnited States
Philadelphia Eye AssociatesPhiladelphia, PennsylvaniaUnited States
Bucci Laser VisionWilkes-Barre, PennsylvaniaUnited States
Vance Thompson VisionSioux Falls, South DakotaUnited States
Baylor College of MedicineHouston, TexasUnited States
Parkhurst NuVisionSan Antonio, TexasUnited States
Hoopes VisionDraper, UtahUnited States
Utah Eye CentersOgden, UtahUnited States
Virginia Eye ConsultantsNorfolk, VirginiaUnited States

+Publications

  • Results publication: Vukich J, Thompson V, Yeu E, Wiley WF, Bafna S, Koch DD, Lin L, Michna M. Evaluating the small aperture intraocular lens: depth of focus and the role of refraction and preoperative corneal astigmatism in visual performance. J Cataract Refract Surg. 2024 Nov 1;50(11):1165-1172. doi: 10.1097/j.jcrs.0000000000001524. PMID 39075732

+Study documents

+Investigators