PRECISION Study: Multi-center Study of Performance of a Novel Implanted CGM System Using a Next Generation Transmitter and Algorithm

NCT03251079 — Not applicable, Completed, 36 participants, Senseonics, Inc.

ClinicalTrials Daily  ·  Areas  ·  Conditions  ·  Drugs  ·  Sponsors  ·  

PRECISION Study: A Prospective, Multi-center Evaluation of Precision, Compression and Performance of a Novel Implanted Continuous Glucose Sensor Using a Next Generation Transmitter and Algorithm

RegistryNCT03251079, acronym PRECISION
StatusCompleted
PhaseNot applicable
Study typeInterventional; na, single group, masking: none, diagnostic
Enrollment36 (Actual)
SponsorSenseonics, Inc. (Industry)
ConditionsDiabetes Mellitus, Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2
InterventionsContinuous glucose monitoring system (Device)
Datesstart 2017-07-25, primary completion 2018-02-01, completion 2018-02-01, first posted 2017-08-16, last updated 2024-02-12
Locations3 sites in United States

Related Papers Related Patents Related Grants Related Experts
AI Assistant Save
All trials for Diabetes Mellitus All trials for Diabetes Mellitus, Type 1 All trials of Continuous glucose monitoring system

-Summary

The purpose of this multi-center, prospective clinical study is to evaluate the performance of a novel, implanted Senseonics continuous glucose monitoring system (Senseonics CGM System) compared to Yellow Springs Instrument (YSI) glucose analyzer reference standard measurements using a next generation transmitter and algorithm. Other measures evaluated will be the effects of compression on performance and the safety of the Senseonics CGM system.

Expand all sections  ·  Collapse all

+Arms and interventions

ArmTypeInterventionsDescription
Continuous Glucose Monitoring DeviceExperimentalDevice: Continuous glucose monitoring systemSenseonics continuous glucose monitoring system
  • Continuous glucose monitoring system (Device), also known as Senseonics continuous glucose monitoring system: Implanted continuous glucose monitoring system

+Outcome measures

Primary outcomes

  • Accuracy of CGM System Through 90 Days Post-Insertion (MARD for Paired CGM and Reference Glucose Measurements) [90 days]: The effectiveness endpoint is the mean absolute relative difference (MARD), calculated for all paired Senseonics CGM System and reference glucose measurements through 90 days of Sensor use. MARD is defined as the average of absolute difference of paired Senseonics CGM System and reference glucose readings divided by the reference glucose reading (reference) for all reference glucose values, that is: MARD = ((SUM \| (Glucose)sensor - (Glucose)reference \| / (Glucose)reference ) / n ) x 100%, where n is the total number of CGM and reference glucose pairs after 90 days of sensor use (MARD is a percentage). Lower MARDs indicate higher (better) accuracy.
  • Safety Endpoint [90 days]: Incidence of device-related or sensor insertion/removal procedure-related serious adverse events through 90 days post-insertion.

+Eligibility

Sex: All, ages: 18 Years to any

Inclusion Criteria:

1. Adult subjects, age ≥18 years
2. Clinically confirmed diagnosis of diabetes mellitus for ≥1 year
3. Subject has signed an informed consent form and is willing to comply with protocol requirements

Exclusion Criteria:

1. History of severe hypoglycemia in the previous 6 months. Severe hypoglycemia is defined as hypoglycemia resulting in loss of consciousness or seizure
2. History of diabetic ketoacidosis requiring emergency room visit or hospitalization in the previous 6 months
3. Female subjects of childbearing capacity (defined as not surgically sterile or not menopausal for ≥ 1 year) who are lactating or pregnant, intending to become pregnant, or not practicing birth control during the course of the study.
4. A condition preventing or complicating the placement, operation, or removal of the Sensor or wearing of transmitter, including upper extremity deformities or skin condition.
5. Symptomatic coronary artery disease; unstable angina; myocardial infarction, transient ischemic attack or stroke in the past 6 months; uncontrolled hypertension (systolic>160 mm Hg or diastolic >100 mm Hg at time of screening); current congestive heart failure; history of cardiac arrhythmia (benign Premature Atrial Contractions (PACs) and Premature Ventricular Contractions (PVCs) allowed).

   Subjects with asymptomatic coronary artery disease (e,g, coronary artery bypass graft (CABG), stent placement or angioplasty) may participate if negative stress test within 1 year prior to screening and written clearance from Cardiologist documented.
6. Hematocrit <30% or >55%
7. History of hepatitis B, hepatitis C, or HIV
8. Current treatment for a seizure disorder unless written clearance by neurologist to participate in study
9. History of adrenal insufficiency
10. Currently receiving (or likely to need during the study period):

    immunosuppressant therapy; chemotherapy; anticoagulant/antithrombotic therapy (excluding aspirin); glucocorticoids (excluding ophthalmic or nasal). This exclusion does include the use of inhaled glucocorticoids and the use of topical glucocorticoids (over sensor site only); antibiotic for chronic infection (e.g. osteomyelitis, endocarditis)
11. A condition requiring or likely to require magnetic resonance imaging (MRI)
12. Known topical or local anesthetic allergy
13. Known allergy to glucocorticoids
14. Any condition that in the investigator's opinion would make the subject unable to complete the study or would make it not in the subject's best interest to participate in the study. Conditions include but are not limited to psychiatric conditions, known current or recent alcohol abuse or drug abuse by subject history, a condition that may increase the risk of induced hypoglycemia or risk related to repeated blood testing. Investigator will supply rationale for exclusion
15. Participation in another clinical investigation (drug or device) within 2 weeks prior to screening or intent to participate during the study period
16. The presence of any other active implanted device (as defined further in protocol)
17. The presence of any other CGM sensor or transmitter located in upper arm (other location is acceptable)

+Posted results

Participant flow

MilestoneContinuous Glucose Monitoring Device
Started36
Completed35
Not completed1

Outcome results

  • Accuracy of CGM System Through 90 Days Post-Insertion (MARD for Paired CGM and Reference Glucose Measurements) (Primary), Mean, percent: Continuous Glucose Monitoring Device 9.6 (8.9 to 10.4)
  • Safety Endpoint (Primary), Number, serious adverse events: Continuous Glucose Monitoring Device 0 (0.0 to 3.5)

Adverse events

Time frame: Through 90 days post Sensor insertion and sensor removal 10-day follow-up (up to 100 days)

ArmDeathsSerious adverse eventsOther adverse events
Continuous Glucose Monitoring Device0/35 (0.0%)3/35 (8.6%)15/35 (42.9%)

Serious adverse events by term

TermOrgan systemContinuous Glucose Monitoring Device
Other - GastroenteritisGastrointestinal disorders1/35
Hypoglycemic episodeEndocrine disorders1/35
Other - Wound infection left foot, grade 4Musculoskeletal and connective tissue disorders1/35

Other adverse events by term

TermOrgan systemContinuous Glucose Monitoring Device
Common ColdRespiratory, thoracic and mediastinal disorders4/35
Sensor Location Site - Pain/DiscomfortSkin and subcutaneous tissue disorders2/35
Other - Upper Respiratory InfectionRespiratory, thoracic and mediastinal disorders3/35
HeadacheNervous system disorders4/35
Other - PneumoniaRespiratory, thoracic and mediastinal disorders1/35
Other - Pilonidal cystSkin and subcutaneous tissue disorders1/35
Other - Difficulty removing sensorSkin and subcutaneous tissue disorders1/35
Adhesive Patch Location Site - Irritation including redness, excoriation or ulcerationSkin and subcutaneous tissue disorders1/35
Other - Darkening of skin at proximal end of sensorSkin and subcutaneous tissue disorders1/35
Other - Vaginal InfectionReproductive system and breast disorders1/35
Other - Dental CariesGeneral disorders1/35
Other - ArthritisMusculoskeletal and connective tissue disorders1/35

+Locations (3)

FacilityCityCountryStatus
AMCR Institute Inc.Escondido, CaliforniaUnited States
Diablo Clinical ResearchWalnut Creek, CaliforniaUnited States
Rainier Clinical Research CenterRenton, WashingtonUnited States