Post-market Clinical Follow-up of a Magnetic Bone Conduction Implant (Cochlear Baha Attract System)

NCT02022085 — Not applicable, Completed, 54 participants, Cochlear Bone Anchored Solutions

ClinicalTrials Daily  ·  Areas  ·  Conditions  ·  Drugs  ·  Sponsors  ·  
RegistryNCT02022085
StatusCompleted
PhaseNot applicable
Study typeInterventional; na, single group, masking: none, treatment
Enrollment54 (Actual)
SponsorCochlear Bone Anchored Solutions (Industry)
ConditionsDeafness, Hearing Loss, Hearing Loss, Conductive
InterventionsBaha Attract System (Device)
Datesstart 2014-06-01, primary completion 2016-02-01, completion 2017-11-01, first posted 2013-12-27, last updated 2018-10-12
Locations6 sites in United States, Netherlands, Poland, United Kingdom

Related Papers Related Patents Related Grants Related Experts
All trials for Deafness All trials for Hearing Loss All trials of Baha Attract System

-Summary

The rationale behind this post-market clinical follow-up investigation is to collect data regarding the usability and clinical performance of the Baha Attract System in subjects with hearing impairment that are candidates for Baha surgery:

* to evaluate the efficacy of the Baha Attract System in terms of hearing performance compared to the unaided situation and compared to a pre-operative test situation using the sound processor on a Baha Softband;
* to evaluate the mid- and long-term safety of the Baha Attract System.

Expand all sections  ·  Collapse all

+Detailed description

This investigation was designed to collect data regarding the usability and clinical performance of the Baha Attract System in subjects with hearing impairment that were candidates for Baha surgery.

The investigation was performed in an open design since it was not possible to perform the investigation in a blinded fashion. The main evaluations of the investigation, i.e. free-field hearing tests, are standard audiological procedures that are routinely used in daily clinical practice at hospitals worldwide for evaluation of hearing performance in hearing impaired patients. The primary efficacy evaluation, audiometric thresholds, renders comparable results across different sites and countries as it is not a language-specific test. Language-specific audiological tests were performed as secondary efficacy evaluations; to enable comparison of data across sites, validated word lists were used at all sites and equivalent speaker configurations (speech from front, noise from behind) and test procedures (same noise and adaptive speech levels, etc) were used.

For the primary evaluation in the investigation, the Baha Attract System was compared to the pre-operative unaided situation, as the intended use of the system is to improve hearing performance in patients with conductive or mixed hearing loss or single-sided sensorineural deafness. Audiological test results with the Baha Attract System were also compared against results obtained with the same Sound Processor on a Baha Softband. Use of the Sound Processor on a Softband is a standard procedure that is routinely used as a pre-operative test prior to bone conduction hearing implant surgery. The Softband test allows the patient to pre-operatively experience hearing through the Sound Processor to get an indication of the post-operative hearing outcome and to choose a suitable Sound Processor.

The generic and hearing-specific quality of life questionnaires used in the investigation (HUI3, APHAB, SSQ) are validated scales that have been frequently used and reported in the literature.

+Arms and interventions

ArmTypeInterventionsDescription
Baha Attract SystemExperimentalDevice: Baha Attract SystemThis transcutaneous solution is based on a magnet coupling using magnets on both side of the skin; * One implant magnet * One external magnet on which a sound processor is attached, i.e. the Sound Processor magnet (SP magnet)
  • Baha Attract System (Device): The Cochlear Baha Attract System incorporates both implantable and external parts. The implantable parts include the BI300 Implant and the BIM400 Implant Magnet, which is intended to be fixated to the BI300 Implant. The external parts include the Sound Processor Magnet (SP Magnet), which, together with the Implant Magnet, constitute the transcutaneous coupling. The SP Magnet incorporates a soft material (Soft pad) at the tissue facing surface that is designed to distribute the pressure over the skin. The Baha Sound Processor attaches to the SP Magnet via a snap coupling.

+Outcome measures

Primary outcomes

  • Hearing Performance: Threshold Audiometry PTA4: Unaided Versus Baha Attract [Baseline (unaided) before surgery, 6, 12 and 24 months after surgery]: The change of hearing performance with the Baha Attract System (aided) at 6, 12 and 24 months from the unaided hearing performance before surgery; measured as free-field hearing tests: Threshold audiometry PTA4 (mean of 500, 1000, 2000 and 4000 Hz). The units reported for PTA4 are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.
  • Hearing Performance: Change in Threshold Audiometry: Unaided Versus Baha Attract [Baseline (unaided) before surgery, 24 months after surgery]: The change of hearing performance with the Baha Attract System (aided) from the unaided hearing performance before surgery; measured as free-field hearing tests: Threshold audiometry PTA4 (mean of 500, 1000, 2000 and 4000 Hz). The units reported for PTA4 are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.
  • Hearing Performance: Change in Threshold Audiometry: Unaided Versus Baha Attract [Baseline (unaided) before surgery, 12 months after surgery]: The change of hearing performance with the Baha Attract System (aided) from the unaided hearing performance before surgery; measured as free-field hearing tests: Threshold audiometry at frequencies of 250, 500, 1000, 2000, 3000, 4000 and 6000 Hz). The units reported for threshold audiometry are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.
  • Hearing Performance: Change in Threshold Audiometry: Unaided Versus Baha Attract [Baseline (unaided) before surgery, 6 months after surgery]: The change of hearing performance with the Baha Attract System (aided) from the unaided hearing performance before surgery; measured as free-field hearing tests: Threshold audiometry at individual frequencies 250, 500, 1000, 2000, 3000, 4000 and 6000 Hz. The units reported for threshold audiometry are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.

Secondary outcomes

  • Adaptive Speech Recognition in Noise Ratio: Unaided Versus Baha Attract [Baseline (unaided) before surgery, 6, 12 and 24 months after surgery]: The change of hearing performance with the Baha Attract System compared to the pre-operative unaided situation; Adaptive speech recognition in noise measured as signal to noise ratio The adaptive speech test in noise was conducted using validated lists of phonetically balanced sentences, with speech presented from the front (0 degrees azimuth) and noise from the back (180 degrees azimuth). The noise was kept constant at 65 dB SPL, and the speech was adapted in 2 dB steps to establish the speech-to-noise ratio (SNR) providing a 50% level of understanding. A ratio of 1 reflects the ability to correctly hear sentences at 65 dB, in the presence of 65 dB background noise. A lower ratio than 1 reflects the ability to correctly hear sentences below 65 dB. A ratio higher than 1 reflects the ability to correctly hear sentences presented above 65 dB. Manchester and Birmingham are excluded from this analysis due to incorrect set up.
  • Speech in Quiet, Baha Attract Versus Unaided [Baseline before surgery, 6, 12 and 24 months after surgery]: The change of hearing performance with the Baha Attract System at 6, 12 and 24 months compared to the pre-operative unaided situation; Speech in quiet at 50, 65 and 80dB. Speech in Quiet is a measure of percentage of correct words. A higher score reflects a higher percentage of correct words.
  • Hearing Performance: Threshold Audiometry PTA4: Sound Processor on Softband Versus Baha Attract [Baseline before surgery, 6, 12 and 24 months after surgery]: The change of hearing performance with the Baha Attract System (aided) from the aided hearing performance, sound processor on a softband, before surgery; measured as free-field hearing tests: Threshold audiometry PTA4 (mean of 500, 1000, 2000 and 4000 Hz) The units reported for PTA4 are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.
  • Hearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband [Baseline (aided) before surgery, 6 months after surgery]: The change of hearing performance with the Baha Attract System (aided) from the aided hearing performance with Sound processor on a Softband before surgery; measured as free-field hearing tests: Threshold audiometry at individual frequencies 250, 500, 1000, 2000, 3000, 4000, 6000 Hz. The units reported for threshold audiometry are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.
  • Hearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband [Baseline (aided) before surgery, 12 months after surgery]: The change of hearing performance with the Baha Attract System (aided) from the aided hearing performance with Sound processor on a Softband before surgery; measured as free-field hearing tests: Threshold audiometry at individual frequencies: 250, 500, 1000, 2000, 3000, 4000, 6000 Hz. The units reported for threshold audiometry are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.
  • Hearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband [Baseline (aided) before surgery, 24 months after surgery]: The change of hearing performance with the Baha Attract System (aided) from the aided hearing performance with Sound processor on a Softband before surgery; measured as free-field hearing tests: Threshold audiometry at individual frequencies: 250, 500, 1000, 2000, 3000, 4000, 6000 Hz. The units reported for threshold audiometry are decibels (dB). As such, a lower or more negative score is more desirable and reflects the ability to hear softer sounds.
  • Adaptive Speech Recognition in Noise Ratio: Baha Attract Versus Sound Processor on Softband [Baseline (aided) before surgery, 6, 12 and 24 months after surgery]: The change of hearing performance with the Baha Attract System compared to the pre-operative aided situation with Sound Processor on a softband; Adaptive Speech recognition in Noise measured as signal to noise ratio. A ratio of 1 reflects the ability to correctly hear sentences at 65 dB, in the presence of 65 dB background noise. A lower ratio than 1 reflects the ability to correctly hear sentences below 65 dB. A ratio higher than 1 reflects the ability to correctly hear sentences presented above 65 dB.
  • Speech in Quiet: Baha Attract Versus Sound Processor on Softband [Baseline (aided) before surgery, 6, 12 and 24 months after surgery]: The change of hearing performance with the Baha Attract System at 6 months compared to the pre-operative aided situation with the Sound Processor on a softband; Speech in quiet at 50, 65 and 80dB. Speech in Quiet is a measure of percentage of correct words. A higher score reflects a higher percentage of correct words.
  • Health Utility Index (HUI) [Baseline (unaided) before surgery, 6 and 24 months after surgery]: Change of health status and health related quality of life using the generic quality of life scale Health Utilities Index (HUI3) when wearing Baha Attract System compared to the pre-operative unaided situation. A health utility value of 1.00 indicates perfect health while a score of 0.00 indicates death. The change from unaided to aided hearing is presented. A positive value indicates an improved quality of life, a negative value indicates impaired quality of life. HUI score of 1 (maximum) describes a state of perfect health. HUI score of 0 describes a state of dead. A negative score of Comprehensive Health State describes a state worse than dead.
  • Abbreviated Profile of Hearing Aid Benefit (APHAB) [Baseline before surgery, 6 and 24 months after surgery]: Measuring change of Ease of communication, Reverberation, Background noise, Aversiveness and a Global score with the Baha Attract System from the pre-operative unaided situation. The absolute APHAB scale is between 0 and 100%, where 0% indicates no problems and 100% indicates always problem. The change from unaided to aided hearing is presented. A positive value indicates an improvement, a negative value an impairment.
  • Speech, Spatial and Qualities of Hearing Scale (SSQ) [Baseline before surgery, 6 and 24 months after surgery]: Measuring change of speech, spatial and hearing experiences with the Baha Attract System from the pre-operative unaided situation. A scale from 0 to 10, where 0 represents "can not hear at all", and 10 "hear perfectly". The change from unaided to aided hearing is presented. A positive value indicates improved hearing, a negative value indicates impaired hearing.
  • Time to Perform Surgery [Visit 2 (Surgery)]: Time of first incision to time of last suture
  • Tissue Reduction Performed During Surgery [Visit 2 (surgery)]: Surgical thinning of the soft tissue flap was advocated when the soft tissue thickness exceeded 6 mm
  • Implant Stability [Visit 2 (surgery)]: Implant Stability Quotient - ISQ, a scale from 1 to 100, where 100 represent the highest stability
  • Magnetic Force [4, 6, 12 weeks, 6, 12 and 24 months]: To investigate if the magnetic force required for sound processor magnet retention will change over time
  • Sound Processor Magnet Choice [4, 6, 12 weeks, 6, 12 and 24 months]: To investigate how sound processor magnet choice will change over time. Six different magnetic strength could be chosen; SPM 1 had the lowest strength and SPM 6 the the highest.
  • Pain & Discomfort [Week 6, Week 12, Month 6, Month 12, Month 24]: Degree of pain and discomfort.
  • Numbness When Tested With a Pin [Day 10, Week 4, Week 6, Week 12, Month 6, Month 12, Month 24]: Degree of numbness when tested with a pin
  • Numbness When Tested With a Cotton Swab [Day 10, Week 4, Week 6, Week 12, Month 6, Month 12, Month 24]: Degree of numbness when tested with a Cotton Swab

+Eligibility

Sex: All, ages: 18 Years to any

Inclusion Criteria:

Adult subject, i.e. ≥ 18 years of age

Conductive or mixed hearing loss in the ear to be implanted:

Bone conduction thresholds with a pure tone average PTA4 of < 30 dB hearing level (mean of 500, 1000, 2000 and 4000 Hz).

OR

Single-sided sensorineural deafness (SSD):

European sites: Bone conduction thresholds with a pure tone average PTA4 of < 30dB hearing level (mean of 500, 1000, 2000 and 4000 Hz) in the good ear.

US sites: Air conduction thresholds with a pure tone average PTA4 of 20 dB hearing level (mean of 500, 1000, 2000 and 3000 Hz) in the good ear OR Subject is indicated for an AC CROS but-for some reason-cannot or will not use an AC CROS.

No previous bone conduction implant on the side of the skull to be implanted. Signed informed consent.

Exclusion Criteria:

Subjects that are scheduled for simultaneous bilateral implant surgery. The investigation is limited to subjects with unilateral use of the Baha Attract System (however, bilateral hearing loss is not an exclusion criterion).

Suitable implant position for the BI300 Implant (4 mm or 3 mm) not found during surgery due to insufficient bone quality and/or bone thickness.

Less than 3 mm soft tissue thickness at the planned implant site. Subjects that have received radiation therapy at the same side of the skull where the Baha Attract System will be positioned.

Condition that could jeopardise osseointegration and/or wound healing as judged by the investigator (e.g. osteoporosis, psoriasis, use of corticosteroids).

Uncontrolled diabetes as judged by the investigator. Condition that may have an impact on the outcome of the investigation as judged by the investigator.

Unable to follow investigational procedures (e.g. to complete quality of life scales).

Participation in another investigation with pharmaceuticals and/or medical device.

+Posted results

Participant flow

MilestoneBaha Attract System
Started54
Completed53
Not completed1

Outcome results

  • Hearing Performance: Threshold Audiometry PTA4: Unaided Versus Baha Attract (Primary), Mean, dB: Baha Attract System -21.0 +/- 10.4 (p <0.0001)
  • Hearing Performance: Change in Threshold Audiometry: Unaided Versus Baha Attract (Primary), Mean, dB: Baha Attract System -19.9 +/- 11.0 (p <0.0001)
  • Hearing Performance: Change in Threshold Audiometry: Unaided Versus Baha Attract (Primary), Mean, dB: Baha Attract System -18.5 +/- 14.3 (p <0.0001)
  • Hearing Performance: Change in Threshold Audiometry: Unaided Versus Baha Attract (Primary), Mean, dB: Baha Attract System -16.9 +/- 12.9 (p <0.0001)
  • Adaptive Speech Recognition in Noise Ratio: Unaided Versus Baha Attract (Secondary), Mean, dB: Baha Attract System -4.30 +/- 5.64 (p <0.0001)
  • Speech in Quiet, Baha Attract Versus Unaided (Secondary), Mean, Change of % correct words at the dB leve: Baha Attract System 52.1 +/- 29.2 (p <0.0001)
  • Hearing Performance: Threshold Audiometry PTA4: Sound Processor on Softband Versus Baha Attract (Secondary), Mean, dB: Baha Attract System -0.62 +/- 5.138 (p 0.38)
  • Hearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband (Secondary), Mean, dB: Baha Attract System -1.30 +/- 9.77 (p 0.34)
  • Hearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband (Secondary), Mean, dB: Baha Attract System -2.55 +/- 11.15 (p 0.11)
  • Hearing Performance: Threshold Audiometry Individual Frequencies: Baha Attract Versus Sound Processor on Softband (Secondary), Mean, dB: Baha Attract System -3.67 +/- 8.35 (p 0.0051)
  • Adaptive Speech Recognition in Noise Ratio: Baha Attract Versus Sound Processor on Softband (Secondary), Mean, change in SNR dB: Baha Attract System -0.508 +/- 4.173 (p 0.46)
  • Speech in Quiet: Baha Attract Versus Sound Processor on Softband (Secondary), Mean, Change of % correct words at the dB leve: Baha Attract System 2.74 +/- 29.94 (p 0.43)
  • Health Utility Index (HUI) (Secondary), Mean, units on a scale: Baha Attract System 0.063 +/- 0.249 (p 0.088)
  • Abbreviated Profile of Hearing Aid Benefit (APHAB) (Secondary), Mean, units on scale: Baha Attract System 20.4 +/- 24.9 (p <0.0001)
  • Speech, Spatial and Qualities of Hearing Scale (SSQ) (Secondary), Mean, units on scale: Baha Attract System 2.49 +/- 1.67 (p <0.0001)
  • Time to Perform Surgery (Secondary), Mean, Minutes: Baha Attract System 38.7 +/- 10.7
  • Tissue Reduction Performed During Surgery (Secondary), Number, participants: Baha Attract System 12
  • Implant Stability (Secondary), Mean, units on a scale: Baha Attract System 77.3 +/- 9.0
  • Magnetic Force (Secondary), Mean, Newton: Baha Attract System 1.01 +/- 0.40
  • Sound Processor Magnet Choice (Secondary), Count of participants, Participants: Baha Attract System 19
  • 3 more outcome measures not shown.

Adverse events

Time frame: 24 months

ArmDeathsSerious adverse eventsOther adverse events
Baha Attract System1/54 (1.9%)13/54 (24.1%)29/54 (53.7%)

Serious adverse events by term

TermOrgan systemBaha Attract System
InfectionInfections and infestations1/54
Increase in instability of knee prosthesisGeneral disorders1/54
PancreatitisGastrointestinal disorders1/54
Embolisation of the Splenic Artery from pancreatitisBlood and lymphatic system disorders1/54
Increase of Meniere AttacksEar and labyrinth disorders1/54
Progression of Hearing LossEar and labyrinth disorders1/54
Decompensated Alcoholic Liver DiseaseHepatobiliary disorders1/54
Broken AnkleInjury, poisoning and procedural complications1/54
CholesteatomaNeoplasms benign, malignant and unspecified (incl cysts and polyps)1/54
Pain/DiscomfortInjury, poisoning and procedural complications1/54
Surgery for Baha Attract on other earSurgical and medical procedures1/54
Increase in Occlusive Arterial DiseaseVascular disorders1/54
DequervianMusculoskeletal and connective tissue disorders1/54
Patient was addmitted to the hospital due to childbirthPregnancy, puerperium and perinatal conditions1/54

Other adverse events by term

TermOrgan systemBaha Attract System
Ear and Labyrinth DisordersEar and labyrinth disorders6/54
Gastrointestinal DisordersGastrointestinal disorders4/54
General Disorders and Administration Site ConditionsGeneral disorders10/54
Infections and InfestationInfections and infestations14/54
Injury, Poisoning and Procedural ComplicationsInfections and infestations5/54
Musculoskeletal and Connective Tissue DisordersMusculoskeletal and connective tissue disorders4/54
Nervous System DisordersNervous system disorders5/54
Respiratory, Thoracic and Mediastinal DisordersRespiratory, thoracic and mediastinal disorders3/54
Skin and Subcutaneous Tissue DisordersSkin and subcutaneous tissue disorders7/54
Vascular DisordersVascular disorders3/54
Predefined event: Pressure related Skin ComplicationsSkin and subcutaneous tissue disorders6/54
Predefined event: Pain/DiscomfortSkin and subcutaneous tissue disorders24/54

+Locations (6)

FacilityCityCountryStatus
Dr. Doug BackhusSeattle, WashingtonUnited States
Christina RungeMilwaukee, WisconsinUnited States
Dr. Myrthe HolNijmegenNetherlands
World Hearing Center Institute of Physiology and Pathology of HearingKajetany, NadarzynPoland
Peter MonksfieldBirminghamUnited Kingdom
Kevin GreenManchesterUnited Kingdom

+Study documents

+Investigators