Safety and Performance Study of the RELIANCE 4-Front Passive Fixation Lead

NCT01856491 — Not applicable, Completed, 167 participants, Boston Scientific Corporation

ClinicalTrials Daily  ·  Areas  ·  Conditions  ·  Drugs  ·  Sponsors  ·  

RELIANCE 4-FRONT™ Passive Fixation Defibrillation Lead Post Market Clinical Follow-Up (PMCF) Study

RegistryNCT01856491
StatusCompleted
PhaseNot applicable
Study typeInterventional; na, single group, masking: none, treatment
Enrollment167 (Actual)
SponsorBoston Scientific Corporation (Industry)
ConditionsTachycardia, Ventricular Fibrillation, Ventricular Flutter
InterventionsRELIANCE 4-FRONT™ Passive Fixation lead implantation (Device)
Datesstart 2013-12-16, primary completion 2015-11-02, completion 2017-11-13, first posted 2013-05-17, last updated 2019-04-30
Locations11 sites in Germany, Israel, Italy

Related Papers Related Patents Related Grants Related Experts
AI Assistant Save
All trials for Tachycardia All trials for Ventricular Fibrillation All trials of RELIANCE 4-FRONT™ Passive Fixation lead implantation

-Summary

The objective of this study is to gather data to establish the chronic safety, performance and effectiveness of the RELIANCE 4-FRONT™ Passive Fixation Defibrillation Leads.

Expand all sections  ·  Collapse all

+Detailed description

The RELIANCE 4-FRONT Passive Fixation Defibrillation Lead Post Market Clinical Follow-Up Study is a prospective, non-randomized, multi-center, single-group, post market clinical study to establish the chronic safety, performance and effectiveness of the RELIANCE 4-FRONT passive fixation defibrillation leads.

A total of 167 patients (including 10 % attrition) are required to evaluate the Primary Endpoint.

Up to 10 Investigational centers located in Europe and Israel. Lead-related Complication-Free Rate (CFR) from Implant through 3 Months Post-Implant.

Lead-related complications associated with the RELIANCE 4-FRONT passive fixation lead will count toward this endpoint.

Lead-related Complication-Free Rate from 3 Months through 15 Months Post-Implant.

Lead-related complications associated with the RELIANCE 4-FRONT passive fixation lead will count toward this endpoint.

* Lead-related Complication-Free Rate from 3 Months through 24 Months Post-Implant
* Pacing Threshold at 0.5 ms pulse width at 3 Months Post-Implant
* Sensed Amplitude at 3 Months Post-Implant
* Pacing Impedance at 3 Months Post-Implant All endpoints will be assessed for the RELIANCE 4-FRONT passive fixation lead.

Clinic visits will occur at:

* Enrollment Visit (no later than 30 days prior to implant procedure)
* Implant Procedure (Day 0; all future follow ups based on this date)
* Pre-Discharge Clinic Visit (3 - 72 hours post-implant)
* One Month Clinic Visit (30±7 days)
* 3 Month Clinic Visit (91 ± 21 days)
* 6 Month Clinic Visit (180 ± 30 days)
* 12 Month Clinic Visit (365 ± 45 days)
* 18 Month Clinic Visit (545± 45 days)
* 24 Month Clinic Visit (730 ± 45 days) The study will be considered completed after all subjects have completed the 24 Month follow-up and study completion is anticipated in 2015. Primary endpoint completion is anticipated after all subjects have completed the 3 Month follow-up visit. All study required visits will be completed during clinic visits.

Hypotheses testing in the RELIANCE 4-FRONT passive fixation PMCF Study will use standard statistical methodology as specified more in detail in the protocol.

+Arms and interventions

ArmTypeInterventionsDescription
RELIANCE 4-FRONT™ Passive FixationOtherDevice: RELIANCE 4-FRONT™ Passive Fixation lead implantationSingle arm, all patients will be implanted with the RELIANCE 4-FRONT™ Passive Fixation lead
  • RELIANCE 4-FRONT™ Passive Fixation lead implantation (Device): Implantation of transvenous defibrillation lead with passive fixation mechanism.

+Outcome measures

Primary outcomes

  • Complication Free Rate [3-months]: Lead-related Complication-Free Rate (CFR) from Implant through 3 Months Post-Implant.

Secondary outcomes

  • Complication Free Rate [3 months through 15 months post implant]: Lead-related Complication-Free Rate from 3 Months through 15 Months Post-Implant.

Other outcomes

  • Pacing Threshold at 0.5 ms Pulse Width [3 Months Post-Implant]: Pacing Threshold at 0.5 ms pulse width at 3 Months Post-Implant. During the RELIANCE 4-FRONT PASSIVE fixation PMCF Study pacing threshold measurements are collected from RELIANCE 4-FRONT PASSIVE fixation leads in the standard manual fashion. At least 3 cardiac cycles at a given voltage level shall be obtained before stepping down to the next voltage level. A count of two non-capture beats is required at a given voltage level to declare a loss of capture (LOC) for any of these tests. The threshold is defined as one voltage level above the level where two non-captured beats are observed. Threshold tests must be taken with a pulse width of 0.5 ms.
  • Complication Free Rate [3 months through 24 months post implant]: Lead-related Complication-Free Rate from 3 Months through 24 Months Post-Implant
  • Sensed Amplitude [3 Months Post-Implant]: Lead performance evaluations of the Reliance 4-FRONT PASSIVE fixation lead have been conducted in accordance with the Physician's Lead Manual. Measurements should fall within the recommended values as mentioned below: Signal Type Amplitude Pacing Threshold Impedance Pacing/Sensing ≥ 5mV ≤1.5 V 300-1200 Ω Defibrillation ≥ 1mV N/A 20-125 Ω The required data from the implant procedure for the study lead is measured with a pacing system analyzer (PSA) to verify adequate signals. Electrical performance of the lead was verified before attaching the lead to the Pulse Generator (PG). After the PG has been implanted evaluation of the study lead using the PG has been performed. Lead measurements were required unless the testing is prohibited by a subject's condition (subject has no intrinsic rhythm). In case the measurements are variable (e.g. in patients with atrial fibrillation), the most reproducible value has been collected.
  • Pacing Impedance [3 Months Post-Implant]: Lead performance evaluations of the Reliance 4-FRONT PASSIVE fixation lead have been conducted in accordance with the Physician's Lead Manual. Measurements should fall within the recommended values as mentioned below: Signal Type Amplitude Pacing Threshold Impedance Pacing/Sensing ≥ 5mV ≤1.5 V 300-1200 Ω Defibrillation ≥ 1mV N/A 20-125 Ω The required data from the implant procedure for the study lead is measured with a pacing system analyzer (PSA) to verify adequate signals. Electrical performance of the lead was verified before attaching the lead to the Pulse Generator (PG). After the PG has been implanted evaluation of the study lead using the PG has been performed. Lead measurements were required unless the testing is prohibited by a subject's condition (subject has no intrinsic rhythm). In case the measurements are variable (e.g. in patients with atrial fibrillation), the most reproducible value has been collected.

+Eligibility

Sex: All, ages: 18 Years to any

Inclusion Criteria:

* Willing and capable of providing informed consent
* Has an indication for implantation of a single or dual chamber ICD or CRT-D system in their respective geography
* Subjects planned to be implanted with the RELIANCE 4-FRONT Passive Fixation Lead
* Willing and capable of participating in all testing/ visits associated with this clinical study at an approved clinical study center and at the intervals defined by this protocol
* Age 18 or above, or of legal age to give informed consent specific to state and national law

Exclusion Criteria:

* Known or suspected sensitivity to Dexamethasone Acetate (DXA)
* Mechanical tricuspid heart valve
* Subject is enrolled in any other concurrent study without prior written approval from Boston Scientific (BSC), with the exception of local mandatory governmental registries and observational studies/registries that are not in conflict and do not affect the following:
* Schedule of procedures for the RELIANCE 4-Front Study (i.e. should not cause additional or missed visits);
* RELIANCE 4-Front Study outcome (i.e. involve medications that could affect the heart rate of the subject);
* Conduct of the RELIANCE 4-Front Study per Good Clinical Practice (GCP)/ International Organization for Standardization (ISO) 14155:2011/ 21 CFR 812/ local regulations
* Currently on the active heart transplant list
* Documented life expectancy of less than 12 months
* Women of childbearing potential who are or might be pregnant at the time of study enrollment (method of assessment upon physician discretion)
* Currently requiring chronic dialysis

+Posted results

Participant flow

MilestoneRELIANCE 4-FRONT™ Passive Fixation
Started167
Completed152
Not completed15

Outcome results

  • Complication Free Rate (Primary), Number, percentage of patients without event: RELIANCE 4-FRONT™ Passive Fixation 97.5 (94.4 to 100)
  • Complication Free Rate (Secondary), Number, percentage of patients without events: RELIANCE 4-FRONT™ Passive Fixation 99.4 (95 to 100)
  • Pacing Threshold at 0.5 ms Pulse Width (Other pre specified), Mean, Volt: RELIANCE 4-FRONT™ Passive Fixation 0.48 +/- 0.59
  • Complication Free Rate (Other pre specified), Number, percentage of patients without events: RELIANCE 4-FRONT™ Passive Fixation 99.4 (95 to 100)
  • Sensed Amplitude (Other pre specified), Mean, mVolt: RELIANCE 4-FRONT™ Passive Fixation 18.5 +/- 5.7
  • Pacing Impedance (Other pre specified), Mean, Ohm: RELIANCE 4-FRONT™ Passive Fixation 474 +/- 106

Adverse events

Time frame: 3-month follow-up data

ArmDeathsSerious adverse eventsOther adverse events
RELIANCE 4-FRONT™ Passive Fixation4/167 (2.4%)84/167 (50.3%)55/167 (32.9%)

Serious adverse events by term

TermOrgan systemRELIANCE 4-FRONT™ Passive Fixation
Inappropriate tachy therapy - SVTCardiac disorders1/167
ErosionSkin and subcutaneous tissue disorders1/167
Infection (> 30 days post-implant)Infections and infestations3/167
Psychological effect due to device therapyPsychiatric disorders1/167
Post-surgical wound discomfortSkin and subcutaneous tissue disorders1/167
Post-surgical infection (<= 30 days post-implant)Infections and infestations2/167
Chest painCardiac disorders1/167
Hematoma - Pocket (<=30 days post-implant)Skin and subcutaneous tissue disorders4/167
Conductor coil fracture - RAGeneral disorders1/167
Elevated threshold - RVGeneral disorders2/167
Dislodgment - Unable to capture - LVGeneral disorders1/167
Pneumothorax - ProcedureInjury, poisoning and procedural complications2/167
Ventricular fibrillation (VF)Cardiac disorders3/167
Ventricular tachycardia (VT)Cardiac disorders3/167
Atrial flutterCardiac disorders3/167
Other SVT (AVRT, AVNRT, EAT etc.)Cardiac disorders1/167
Chest pain - Heart failureCardiac disorders3/167
Dyspnea - Heart failureCardiac disorders8/167
Peripheral edema - Heart failureCardiac disorders1/167
Pulmonary edema - Heart failureCardiac disorders4/167
Heart failure symptoms - UnspecifiedCardiac disorders6/167
Multiple heart failure symptomsCardiac disorders4/167
Multi-system failure - Heart failureCardiac disorders5/167
HypotensionVascular disorders1/167
HypertensionVascular disorders1/167
RELATED TO ISCHEMIACardiac disorders1/167
Myocardial infarctionCardiac disorders6/167
Coronary Artery DiseaseVascular disorders2/167
Coronary artery diseaseVascular disorders1/167
Peripheral vascular diseaseVascular disorders3/167
Intermittent claudicationVascular disorders1/167
Mitral regurgitationCardiac disorders1/167
SyncopeCardiac disorders3/167
DizzinessCardiac disorders2/167
Chest pain - IschemicCardiac disorders1/167
Chest pain - OtherCardiac disorders6/167
FatigueGeneral disorders1/167
Cerebrovascular accident (CVA)Vascular disorders6/167
Pulmonary embolism (PE)Vascular disorders1/167
Intracardiac thrombusVascular disorders1/167
Hematoma - Unrelated to procedure or deviceGeneral disorders1/167
DeathGeneral disorders4/167
Systemic infectionInfections and infestations2/167
FeverGeneral disorders2/167
Physical traumaInjury, poisoning and procedural complications2/167
Abnormal laboratory valuesBlood and lymphatic system disorders3/167
HematologicalBlood and lymphatic system disorders5/167
NeurologicalNervous system disorders2/167
GastrointestinalGastrointestinal disorders10/167
PulmonaryRespiratory, thoracic and mediastinal disorders9/167
GenitourinaryRenal and urinary disorders7/167
RenalRenal and urinary disorders7/167
MusculoskeletalMusculoskeletal and connective tissue disorders1/167
IntegumentarySkin and subcutaneous tissue disorders1/167
Head, eyes, ears, nose, throat (HEENT)General disorders2/167
EndocrineEndocrine disorders2/167
Multi-system failureGeneral disorders1/167
sleep apneaRespiratory, thoracic and mediastinal disorders1/167

Other adverse events by term

TermOrgan systemRELIANCE 4-FRONT™ Passive Fixation
Not yet classifiedGeneral disorders3/167
Inappropriate tachy therapy - SVTCardiac disorders4/167
Inappropriate tachy therapy - OtherCardiac disorders1/167
Post-surgical wound discomfortSurgical and medical procedures1/167
Inadvertent VT/VFCardiac disorders2/167
Hematoma - Pocket (<=30 days post-implant)Injury, poisoning and procedural complications2/167
Elevated threshold - RACardiac disorders1/167
Elevated threshold - RVCardiac disorders1/167
Coronary venous dissectionVascular disorders1/167
Ventricular fibrillation (VF)Cardiac disorders2/167
Ventricular tachycardia (VT)Cardiac disorders1/167
Atrial fibrillation (AF)Cardiac disorders4/167
Other SVT (AVRT, AVNRT, EAT etc.)Cardiac disorders3/167
Syncope - Heart failureCardiac disorders1/167
Chest pain - Heart failureCardiac disorders1/167
Dyspnea - Heart failureCardiac disorders7/167
Multiple heart failure symptomsCardiac disorders1/167
Hypertension - Heart failureCardiac disorders1/167
Chest pain - IschemicCardiac disorders1/167
Chest pain - OtherCardiac disorders7/167
PalpitationsCardiac disorders3/167
FatigueGeneral disorders1/167
HemorrhageSurgical and medical procedures1/167
Physical traumaGeneral disorders6/167
HematologicalBlood and lymphatic system disorders1/167
GastrointestinalGastrointestinal disorders5/167
PulmonaryRespiratory, thoracic and mediastinal disorders2/167
GenitourinaryRenal and urinary disorders4/167
RenalRenal and urinary disorders1/167
MusculoskeletalMusculoskeletal and connective tissue disorders6/167
PsychologicalPsychiatric disorders1/167
IntegumentarySkin and subcutaneous tissue disorders3/167
Head, eyes, ears, nose, throat (HEENT)General disorders10/167
EndocrineEndocrine disorders2/167
CancerGeneral disorders2/167
medication side effectGeneral disorders1/167
pain in device pocketSkin and subcutaneous tissue disorders1/167

Limitations and caveats: Sample size is rather small and the power of the study is only 80%

+Locations (11)

FacilityCityCountryStatus
German Heart CenterBerlinGermany
Meir Medical CenterKfar SabaIsrael
Beilinson Medical CenterPetah TikvaIsrael
Sheba Medical CenterTel LitwinskyIsrael
Fondazione di Ricerca e Cura 'Giovanni Paolo IICampobassoItaly
Ospedale Alessandro ManzoniLeccoItaly
Clinica MediterraneaNaplesItaly
Ospedale Buon ConsiglioNaplesItaly
Azienda Ospedaliero Universitaria PisanaPisaItaly
Osp. Civile S. Maria Delle GraziePozzuoliItaly
Ospedale San Pietro FatebenefratelliRomeItaly

+Investigators

My Notes