Data Collection of Arterial Line Waveform and BP Values

NCT04510753 — Completed, 100 participants, Sensifree Ltd.

ClinicalTrials Daily  ·  Areas  ·  Conditions  ·  Drugs  ·  Sponsors  ·  

Evaluation of Sensifree's Algorithm for Continuous Blood Pressure Measurements Based on a PPG Sensor's Signal Compared to Invasive Radial Arterial Line

RegistryNCT04510753
StatusCompleted
Study typeObservational
Enrollment100 (Actual)
SponsorSensifree Ltd. (Industry)
ConditionsBlood Pressure Measurement
InterventionsArterial catheterization kit, GE Datex-Ohmeda Finger Clip Pulse Oximeter Sensor, NIBP Cuff (GE Healthcare) (Device)
Datesstart 2019-11-22, primary completion 2020-12-09, completion 2020-12-09, first posted 2020-08-12, last updated 2022-06-07
Locations1 site in United States

Related Papers Related Patents Related Grants Related Experts
All trials for Blood Pressure Measurement All trials of Arterial catheterization kit All trials of GE Datex-Ohmeda Finger Clip Pulse Oximeter Sensor

-Summary

Collect continuous systolic and diastolic blood pressure values and waveforms as recorded by an arterial line catheter and by non-invasive blood pressure (NIBP) and to collect Photoplethysmogram Pulse Oximetry (PPG) waveform in order to optimize the Sensifree algorithm.

Expand all sections  ·  Collapse all

+Detailed description

Sensifree is developing a continuous non-invasive blood pressure measurement system, that uses a fingertip PPG sensor as a signal source. The system provides systolic, diastolic and mean arterial pressure values, for use in hospitals and other clinical settings. The PPG waveform is analyzed by a proprietary software algorithm, calculating BP values.

The current best method to obtain similar information is by using an Arterial line catheter which will serve as a reference device. In this study, the sponsor aims to prospectively collect continuous radial arterial line \& PPG waveforms, as well as NIBP data over the period of the surgery in order to test and refine the algorithm.

+Arms and interventions

  • Arterial catheterization kit (Device): Invasive monitoring of blood pressure via catheterization of the radial artery, displaying a continuous pressure waveform and blood pressure values.
  • GE Datex-Ohmeda Finger Clip Pulse Oximeter Sensor (Device): PPG fingertip mounted sensor, that measures the oxygen level (oxygen saturation) of the blood and generates a waveform
  • NIBP Cuff (GE Healthcare) (Device): A blood pressure cuff is used to measure blood pressure. The cuff has an inflatable rubber bladder that is typically fastened around the arm. A pressure meter indicates the cuff's pressure.

+Outcome measures

Primary outcomes

  • Comparison of BP values between a PPG sensor based device and an arterial line [The length of each procedure, typically up to 8 hours]: PPG signal will be acquired from the PPG sensor, while blood pressure values (systolic, diastolic, mean) will be acquired from A-line. PPG signal will be analyzed by the Sensifree's algorithm and compared to the A-line BP measurement results according to ISO 81060-2 accuracy specifications

+Eligibility

Sex: All, ages: 18 Years to 80 Years

Inclusion Criteria:

1. Signed inform consent
2. Age 18 or greater
3. Elective surgery that requires invasive BP monitoring (A-line)

Exclusion Criteria:

1. Emergency surgery
2. A-line not inserted
3. Informed consent not signed

+Locations (1)

FacilityCityCountryStatus
San Francisco Veterans Affairs Medical CenterSan Francisco, CaliforniaUnited States

+Investigators