Non-invasive Assessment of Cardiac Output and Fluid Responsiveness in Bariatric Surgery

NCT03184272 — Not applicable, Completed, 50 participants, University Hospital Schleswig-Holstein

ClinicalTrials Daily  ·  Areas  ·  Conditions  ·  Drugs  ·  Sponsors  ·  

Accuracy of Non-invasive Assessment of Cardiac Output and Fluid Responsiveness in Patients Undergoing Bariatric Procedures.

RegistryNCT03184272
StatusCompleted
PhaseNot applicable
Study typeInterventional; na, single group, masking: none, diagnostic
Enrollment50 (Actual)
SponsorUniversity Hospital Schleswig-Holstein (Other)
ConditionsBariatric Surgery, Hemodynamics
Interventionsbaseline alert 1, Anti-Trendelenburg-position (ATB), ATB in narcosis, baseline in narcosis 1, passive leg raising, volume bolus substitution, baseline in narcosis 2, start capnoperitoneum, ATB plus capnoperitoneum, ATB plus capnoperitoneum plus volume bolus substitution, ATB loss of capnoperitoneum, baseline, baseline alert 2, torso position rising 30° at the beginning, torso position rising 30° at the end (Procedure)
Datesstart 2015-06-01, primary completion 2017-06-01, completion 2017-06-01, first posted 2017-06-12, last updated 2017-06-12
Locations1 site in Germany

Related Papers Related Patents Related Grants Related Experts
AI Assistant Save
All trials for Bariatric Surgery All trials for Hemodynamics All trials of baseline alert 1 All trials of Anti-Trendelenburg-position (ATB)

-Summary

Comparability of discontinuous non-invasive (sphygomanometric) and continuous semi-invasive and invasive (Masimo©; Nexfin© Monitoring ; FloTrac © Edwards Lifesciences) beat to beat measurement methods for the determination of cardiac output and fluid responsiveness in patients undergoing bariatric surgery.

Expand all sections  ·  Collapse all

+Detailed description

Comparability of discontinuous non-invasive (sphygomanometric) and continuous semi-invasive and invasive (Masimo©; Nexfin© Monitoring ; FloTrac © Edwards Lifesciences) beat to beat measurement methods for the determination of cardiac output and fluid responsiveness in patients undergoing bariatric surgery.

Bariatric patients show limitations regarding the conventional monitoring options with ECG derivation, pulse oxymetric oxygen saturation and sphygmo-manometric blood pressure measurement. From a clinical point of view this results in an invasive arterial blood pressure monitoring. In addition to that, intraoperative extreme changes of the patients position on the table combined with the applied pneumoperitoneum during the minimally invasive laparoscopic surgery can significantly influence the cardiovascular parameters. An invasive arterial blood pressure monitoring is able to reproduce the blood pressure by stroke, but it does not give any information about the cardiac output. The sphygmomanometric blood pressure measurement also entails the risk of insufficient detection of a hypotonic phase in the measurement. Additional monitoring systems are currently available, which are capable of measuring completely non-invasive or semi-invasively different cardiovascular parameters such as cardiac output (CO) and volume responsiveness. In this study, the investigators will compare continuous and discontinuous cardiovascular monitoring procedures and their parameters. The measurements take place at specific times, under defined changes in the body position with the additional influence of the pneumoperitoneum.

These changes are recorded and compared at the same time during different measuring methods preoperatively, intraoperatively and postoperatively. The non-invasive Nexfin © Monitoring (Edwards Lifesciences) is to be evaluated and compared with further measurement methods (FloTrac © (Edwards Lifesciences)) as well as the sphygmomanometric upper arm blood pressure measurement. The investigators also compare the non-invasive and invasive continuous beat to beat blood pressure measurement with conventional, discontinuous sphygomanometric upper arm blood pressure measurement.

The planned study is the comparison of 3 different hemodynamic monitoring procedures. For the evaluation we use the method used by Bland and Altmann for calculating the mean deviation (bias) and the precision (mean value ± 2 standard deviations. In the case of multiple measurements , the modification of the Bland-Altman method is applied (repeated measurements). The number of cases was determined with n = 60 patients, followed by an intermediate evaluation. For a Bland-Altman analysis, the width w of the confidence interval for the Limits of Agreement is calculated as w = 6.79 • σ • 1 / √n, where n is number of cases and σ is the standard deviation. For a case count of n = 60, the result is w = 0.88 • σ and thus for this explorative study a sufficiently large number. In case of the dynamic variables, the percentage matching and the calculation of the kappa index are also used for the statistical evaluation.

+Arms and interventions

ArmTypeInterventionsDescription
bariatric surgeryExperimentalProcedure: baseline alert 1; Procedure: Anti-Trendelenburg-position (ATB); Procedure: ATB in narcosis; Procedure: baseline in narcosis 1; Procedure: passive leg raising; Procedure: volume bolus substitution; Procedure: baseline in narcosis 2; Procedure: start capnoperitoneum; Procedure: ATB plus capnoperitoneum; Procedure: ATB plus capnoperitoneum plus volume bolus substitution; Procedure: ATB loss of capnoperitoneum; Procedure: baseline; Procedure: baseline alert 2; Procedure: torso position rising 30° at the beginning; Procedure: torso position rising 30° at the endbaseline alert 1; Anti-Trendelenburg-position (ATB); Anti-Trendelenburg-position (ATB) in narcosis; baseline in narcosis 1; passive leg raising; volume bolus substitution (15 ml/kgKG Sterofundin balanced solution); baseline in narcosis 2; start capnoperitoneum; Anti-Trendelenburg-position (ATB) plus capnoperitoneum; ATB plus capnoperitoneum plus volume bolus substitution (15 ml/kgKG Sterofundin balanced solution); ATB loss of capnoperitoneum; baseline in narcosis; baseline alert 2; torso position rising 30° at the beginning; torso position rising 30° at the end
  • baseline alert 1 (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement in supine position
  • Anti-Trendelenburg-position (ATB) (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement under ATB
  • ATB in narcosis (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement under ATB in narcosis
  • baseline in narcosis 1 (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement in supine position in narcosis
  • passive leg raising (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement under passive leg raising
  • volume bolus substitution (Procedure), also known as Device: measurement in supine position after volume bolus (15 ml/kgKG Sterofundin balanced solution intravenous)
  • baseline in narcosis 2 (Procedure), also known as Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement in supine position
  • start capnoperitoneum (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement in supine position
  • ATB plus capnoperitoneum (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement under ATB plus capnoperitoneum
  • ATB plus capnoperitoneum plus volume bolus substitution (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement in ATB plus capnoperitoneum after volume bolus (15 ml/kgKG Sterofundin balanced solution intravenous)
  • ATB loss of capnoperitoneum (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement under ATB
  • baseline (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement in supine position
  • baseline alert 2 (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement in supine position
  • torso position rising 30° at the beginning (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement in the recovery unit
  • torso position rising 30° at the end (Procedure), also known as Device, Nexfin© vs FloTrac© vs sphygmomanometric blood pressure: measurement in the recovery unit

+Outcome measures

Primary outcomes

  • Non-invasive cardiac output, i.e. stroke volume measurement using Nexfin technology [From the beginning of surgery to admission to the PACU, approximately 4 hours]: Accuracy of non-invasive assessment of cardiac output, i.e. stroke volume index in comparison to an invasive standard using FloTrac Vigileo technology

Secondary outcomes

  • Non-invasive assessment of fluid responsiveness using Nexfin technology [From the beginning of surgery to admission to the PACU, approximately 4 hours]: Accuracy of non-invasive assessment of fluid responsiveness (PPV, SVV) in comparison to an invasive standard using FloTrac Vigileo technology

+Eligibility

Sex: All, ages: 18 Years to any

Inclusion Criteria:

* Indication for bariatric surgery was made
* Classification according to the American Society of Anesthesiologists (ASA) 2-4
* Age > 18 years
* Elective laparoscopic surgery
* Written declaration of consent
* Body Mass Index > 30 kg / m²

Exclusion Criteria:

* atrial fibrillation
* cardiac arrhythmias
* aortic aneurysm > 4,5 cm
* Peripheral arterial occlusive desease grade 3-4
* age < 18 years
* missing or incorrect patient consent form
* cognitive or linguistic barriers

+Locations (1)

FacilityCityCountryStatus
Universitätsklinikum Schleswig-Holstein Campus Kiel - Klinik für Anästhesiologie und Operative IntensivmedizinKiel, Deutschland (deu)Germany

+Publications

  • Results publication: Lorenzen U, Pohlmann M, Hansen J, Klose P, Gruenewald M, Renner J, Elke G. Perioperative non-invasive versus semi-invasive cardiac index monitoring in patients with bariatric surgery - a prospective observational study. BMC Anesthesiol. 2020 Aug 10;20(1):196. doi: 10.1186/s12871-020-01110-x. PMID 32778047