Validation of Next Generation INVOS NIRS Cerebral and Tissue Oximeter to Measure Cerebral and Somatic Tissue Oxygen Saturation in Healthy Volunteers
| Registry | NCT03128372 |
| Status | Completed |
| Phase | Not applicable |
| Study type | Interventional; na, single group, masking: none, diagnostic |
| Enrollment | 32 (Actual) |
| Sponsor | Medtronic - MITG (Industry) |
| Conditions | Oxygen Deficiency |
| Interventions | Desaturation (Device) |
| Dates | start 2017-05-22, primary completion 2017-07-19, completion 2017-07-19, first posted 2017-04-25, last updated 2018-11-09 |
| Locations | 1 site in United States |
Related Papers Related Patents Related Grants Related Experts
AI Assistant Save
All trials for Oxygen Deficiency All trials of Desaturation
-Summary
This is a validation study of a near-infrared spectroscopy (NIRS) device designed to measure the cerebral tissue oxygen saturation.
Expand all sections · Collapse all
+Arms and interventions
| Arm | Type | Interventions | Description |
|---|
| Desaturation | Experimental | Device: Desaturation | Volunteers undergo oxygen desaturation in order to determine the accuracy of the device over a clinical range of oxygen saturations 70 - 100%. |
- Desaturation (Device), also known as Cerebral and Tissue Oximeter: The delivered gas mixture will be adjusted to decrease the displayed saturation of peripheral oxygen. Each desaturation steps are of approximately 5 minutes duration with reduction in pulse oximeter oxygen saturation from 100 to 70%.
+Outcome measures
Primary outcomes
- Validated the Next Generation Oximeter [Data collected from individual participant over 4 hour timeframe.]: Cerebral overall mean bias (percentage saturation) defined as the average of the differences between the regional saturation (rSO2) value and (fSO2) value obtained from simultaneous arterial and jugular venous blood samples. Cerebral trending is defined as the measurement of changes in regional saturation (rSO2) under conditions of changing fSO2. Cerebral trending mean bias is the average difference between changes in rSO2 values compared against changes in fSO2. Somatic trending mean bias is defined the same as Cerebral trending mean bias except for the location of measurement on the subjects and for Somatic trending mean bias, rS02 was compared to rSO2 on a commercially-available regional oximetry monitor. The smaller value the better performance.
+Eligibility
Sex: All, ages: 18 Years to 46 Years, accepts healthy volunteers
Inclusion Criteria:
1. Healthy, male or female subjects between the ages of 18 to ≤46 years;
2. Completion of a health screening for a medical history by a licensed physician, nurse practitioner or physician assistant;
3. Minimum weight 40kg;
4. BMI within range 18.0 - 30.0.
Exclusion Criteria:
1. Prior or known allergies to lidocaine (or similar pharmacologic agents, e.g., Novocain) \[self-reported\];
2. Prior known severe allergies to medical grade adhesive/tape (Band-Aid) \[self-reported\];
3. Taking any medication other than birth control \[self-reported\];
4. Is currently participating in, or has recently participated in (discontinued within 30 days prior to the hypoxia procedure for this study) in an investigational drug, device, or biologic study \[self-reported\];
5. Has a negative Allen's Test to confirm non-patency of the collateral artery \[clinical assessment by PI or delegate\];
6. Has made a whole blood donation or has had at least 450 ml of blood drawn within 8 weeks prior to the study procedure \[self-reported\];
7. Is female with a positive pregnancy test \[serum or urine\], or is female and is unwilling to use effective birth control between the time of screening and study procedure or is breastfeeding;
8. Has anemia \[lab values specific for gender\];
9. Has a history of sickle cell trait or thalassemia \[self-reported\];
10. Has an abnormal hemoglobin electrophoresis test \[lab measurement\];
11. Has a positive urine cotinine test or urine drug screen or oral ethanol test \[POC testing\];
12. Has a room air saturation less than 95% by pulse oximetry \[measurement by PI or delegate\];
13. Has a clinically significant abnormal ECG \[assessment by PI or delegate\];
14. Has a clinically significant abnormal pulmonary function test via spirometry \[assessment by PI or delegate\];
15. Has a COHb greater than 3%, or MetHb greater than 2% \[measured by venous blood sample co-oximetry\]
+Posted results
Participant flow
| Milestone | Desaturation |
|---|
| Started | 32 |
| Completed | 32 |
| Not completed | 0 |
Outcome results
- Validated the Next Generation Oximeter (Primary), Mean, Percentage saturation: Desaturation 1.22 +/- 5.92
Adverse events
Time frame: 48 hours post study procedure (desaturation)
| Arm | Deaths | Serious adverse events | Other adverse events |
|---|
| Desaturation | 0/32 (0.0%) | 0/32 (0.0%) | 2/32 (6.3%) |
Other adverse events by term
| Term | Organ system | Desaturation |
|---|
| Vasovagal episode | Nervous system disorders | 1/32 |
| Clavicular pain | Musculoskeletal and connective tissue disorders | 1/32 |
+Locations (1)
| Facility | City | Country | Status |
|---|
| Duke University Hospital, HPPL | Durham, North Carolina | United States | |
+Study documents
+Investigators