| Registry | NCT03103503 |
| Status | Terminated - Collection of long term lead performance is being transferred to an alternate data collection methodology. |
| Study type | Observational |
| Enrollment | 901 (Actual) |
| Sponsor | Biotronik, Inc. (Industry) |
| Conditions | Implantable Defibrillator User |
| Dates | start 2017-05-18, primary completion 2019-09-27, completion 2019-09-27, first posted 2017-04-06, last updated 2020-10-14 |
| Locations | 28 sites in United States |
Related Papers Related Patents Related Grants Related Experts
All trials for Implantable Defibrillator User
-Summary
The purpose of this registry is to confirm the long-term safety and reliability of BIOTRONIK's Plexa ICD lead.
Expand all sections · Collapse all
+Outcome measures
Primary outcomes
- Percentage of Subjects Free From Plexa Lead Related Adverse Events [Up to 2.5 years]: Evaluate the percentage of subjects without an adverse event related to the Plexa lead through study termination. The endpoint is analyzed as the adverse event free-rate. All protocol defined adverse events that occurred more than 30 days after implant and determined to be related to the Plexa lead plus required invasive intervention to resolve or were related to a lead integrity or clinical performance issue were included. Events meeting this criteria but occurring within 30 days were also included except for lead dislodgement, high pacing threshold, failure to capture, or intermittent capture.
Secondary outcomes
- Plexa Lead Safety-Individual Adverse Events [Up to 2.5 years]: Evaluate the individual types of adverse events contributing to primary outcome measure 'Percentage of Subjects Free From Plexa Lead Related Adverse Events'
- Pacing Threshold Measurements for the Plexa Lead Through Study Termination [Up to 2.5 years]: Pacing threshold measurements at pulse width of 0.4 for the Plexa leads at completed follow-up visits.
- Sensing Measurements for the Plexa Lead Through Study Termination [Up to 2.5 years]: Sensing measurements for the Plexa leads at completed follow-up visits.
- Pacing Impedance Measurements for the Plexa Lead Through Study Termination [Up to 2.5 years]: Pacing impedance measurements for the Plexa leads at completed follow-up visits.
- Shock Impedance for the Plexa Lead Through Study Termination [Up to 2.5 years]: Shock impedance measurements for the Plexa leads at completed follow-up visits.
- Percentage of Subjects Free From Protocol Defined Adverse Events [Up to 2.5 years]: Evaluate the percentage of subjects without a protocol defined and clinical event committee adjudicated adverse event, excluding events included in the primary outcome 'Percentage of Subjects Free From Plexa Lead Related Adverse Events', through study termination. Lead dislodgements, high pacing threshold, failure to capture, and intermittent capture related to the Plexa lead that occurred less than or equal to 30 days after lead implant or a lead revision procedure are included here. The endpoint is analyzed as the adverse event free-rate.
+Eligibility
Sex: All, ages: 18 Years to any
Inclusion Criteria:
* Successfully implanted with a BIOTRONIK ICD or CRT-D compatible system along with the Plexa lead no more than 30 days prior to consent
* Able to understand the nature of the registry and provide informed consent
* Available for follow-up visits on a regular basis at the study site for the expected 5 years of follow-up
* Accepts BIOTRONIK Home Monitoring® concept
* Age greater than or equal to 18 years
Exclusion Criteria:
* Enrolled in any investigational cardiac device trial
* Enrolled in BIOTRONIK's QP ExCELs lead study
* Planned cardiac surgical procedures or interventional measures within the next 6 months
* Expected to receive heart transplantation or ventricular assist device within 1 year
* Life expectancy of less than 1 year
* Patients reporting pregnancy at the time of enrollment
+Posted results
Participant flow
| Milestone | BIOTRONIK Plexa ICD Lead |
|---|
| Started | 901 |
| Completed | 0 |
| Not completed | 901 |
Outcome results
- Percentage of Subjects Free From Plexa Lead Related Adverse Events (Primary), Number, percentage of participants: BIOTRONIK Plexa ICD Lead 100.00 (99.49 to 100.00)
- Plexa Lead Safety-Individual Adverse Events (Secondary)
- Pacing Threshold Measurements for the Plexa Lead Through Study Termination (Secondary), Mean, V: BIOTRONIK Plexa ICD Lead 0.63 +/- 0.28
- Sensing Measurements for the Plexa Lead Through Study Termination (Secondary), Mean, mV: BIOTRONIK Plexa ICD Lead 15.88 +/- 5.56
- Pacing Impedance Measurements for the Plexa Lead Through Study Termination (Secondary), Mean, ohms: BIOTRONIK Plexa ICD Lead 567.1 +/- 111.2
- Shock Impedance for the Plexa Lead Through Study Termination (Secondary), Mean, ohms: BIOTRONIK Plexa ICD Lead 75.1 +/- 13.7
- Percentage of Subjects Free From Protocol Defined Adverse Events (Secondary), Number, percentage of participants: BIOTRONIK Plexa ICD Lead 95.12 (93.49 to 96.36)
Adverse events
Time frame: Though study exit or study termination, an average of 1.1 years
| Arm | Deaths | Serious adverse events | Other adverse events |
|---|
| BIOTRONIK Plexa ICD Lead | 64/901 (7.1%) | 44/901 (4.9%) | 12/901 (1.3%) |
Serious adverse events by term
| Term | Organ system | BIOTRONIK Plexa ICD Lead |
|---|
| Hematoma | Surgical and medical procedures | 3/901 |
| Infection | Infections and infestations | 11/901 |
| Loose set-screw | Surgical and medical procedures | 3/901 |
| Non-healing pocket dehiscence requiring intervention | Surgical and medical procedures | 1/901 |
| Cardiogenic shock during implantation | Surgical and medical procedures | 1/901 |
| Operation wound bleeding | Surgical and medical procedures | 1/901 |
| Venous occlusion associated with the implant procedure | Surgical and medical procedures | 1/901 |
| LV lead related high pacing threshold | Cardiac disorders | 1/901 |
| LV lead related lead dislodgement | Cardiac disorders | 1/901 |
| Twiddler's syndrome | General disorders | 2/901 |
| RV lead related lead dislodgement | Cardiac disorders | 12/901 |
| RV lead related other: high defibrillation threshold.no evidence for lead malfunction | Cardiac disorders | 1/901 |
| RA lead related lead dislodgement | Cardiac disorders | 5/901 |
| Pulmonary Emboli | General disorders | 1/901 |
| RV lead related high pacing threshold | Cardiac disorders | 1/901 |
| RV lead related intermittent capture | Cardiac disorders | 1/901 |
| RV lead related lead impedance out of range, high | Cardiac disorders | 1/901 |
| RV lead related lead undersensing | Cardiac disorders | 1/901 |
Other adverse events by term
| Term | Organ system | BIOTRONIK Plexa ICD Lead |
|---|
| Hematoma | Surgical and medical procedures | 2/901 |
| Non-healing pocket dehiscence requiring intervention | Surgical and medical procedures | 1/901 |
| Slow wound healing, no sign for infection | Surgical and medical procedures | 1/901 |
| Thrombosis | Surgical and medical procedures | 1/901 |
| LV lead related extracardiac stimulation | Cardiac disorders | 1/901 |
| LV lead related lead impedance out of range, high | Cardiac disorders | 1/901 |
| RV lead related lead dislodgement | Cardiac disorders | 1/901 |
| RV lead related other: positional atrial undersensing | Cardiac disorders | 1/901 |
| Skin erosion | Skin and subcutaneous tissue disorders | 1/901 |
| RA lead related lead dislodgement | Cardiac disorders | 1/901 |
| RA lead related no capture | Cardiac disorders | 2/901 |
+Locations (28)
| Facility | City | Country | Status |
|---|
| Thomas Hospital Research Office | Fairhope, Alabama | United States | |
| Jeffrey S Goodman MD | Los Angeles, California | United States | |
| Eisenhower Desert Cardiology | Rancho Mirage, California | United States | |
| Florida Hospital | Orlando, Florida | United States | |
| Advent Health Tampa | Tampa, Florida | United States | |
| St. Louis Cardiology Consultants | Alton, Illinois | United States | |
| University of Iowa Hospitals and Clinics | Iowa City, Iowa | United States | |
| Heartland Cardiology | Wichita, Kansas | United States | |
| Research Integrity LLC | Owensboro, Kentucky | United States | |
| Northern Light Cardiology | Bangor, Maine | United States | |
| University of Michigan | Ann Arbor, Michigan | United States | |
| Metro Health Hospital | Wyoming, Michigan | United States | |
| Michigan Heart | Ypsilanti, Michigan | United States | |
| Mercy Clinic Cardiology | Springfield, Missouri | United States | |
| St. Louis Heart and Vascular | St Louis, Missouri | United States | |
| Gateway Cardiovascular Research Center | St Louis, Missouri | United States | |
| Mercy Clinic Heart and Vascular | St Louis, Missouri | United States | |
| Glacier View Research Institute Cardiology | Kalispell, Montana | United States | |
| Deborah Heart and Lung Center | Browns Mills, New Jersey | United States | |
| Novant Health | Winston-Salem, North Carolina | United States | |
| Sanford Medical Center | Fargo, North Dakota | United States | |
| The Ohio State Wexner Medical Center | Columbus, Ohio | United States | |
| Toledo Cardiology | Toledo, Ohio | United States | |
| The Arrhythmia Institute | Newtown, Pennsylvania | United States | |
| Upstate Cardiology | Greenville, South Carolina | United States | |
| Carolina Heart Specialists | Lancaster, South Carolina | United States | |
| Carolina Cardiology Associates | Rock Hill, South Carolina | United States | |
| Cardiovascular Associates Ltd | Virginia Beach, Virginia | United States | |
+Study documents