Evaluation of a Novel Electronic Urine Output Monitor (eUOM)

NCT02195713 — Completed, 20 participants, Potrero Medical

ClinicalTrials Daily  ·  Areas  ·  Conditions  ·  Drugs  ·  Sponsors  ·  
RegistryNCT02195713
StatusCompleted
Study typeObservational
Enrollment20 (Actual)
SponsorPotrero Medical (Industry)
ConditionsOliguria, Polyuria
InterventionsAccuryn (Device)
Datesstart 2014-07-01, primary completion 2015-02-01, completion 2015-02-01, first posted 2014-07-21, last updated 2016-02-22
Locations2 sites in United States

Related Papers Related Patents Related Grants Related Experts
All trials for Oliguria All trials for Polyuria All trials of Accuryn

-Summary

Urine output and urine drain line pressure were monitored while urine was drained into either:

1. Accuryn Urine Output Monitor (Potrero Medical) OR
2. Criticore Monitor (Bard Medical)

Expand all sections  ·  Collapse all

+Detailed description

The purpose of this study is to determine the characteristics of urine output and drain line pressure as measured with the Accuryn urinary output monitoring system and urine output as recorded by a commercially available system (Criticore, Bard Medical) when used with a standard Foley catheter and standard urinary drainage tube.

+Arms and interventions

ArmTypeInterventionsDescription
Test- Accuryn Urine Output MonitorDevice: AccurynAccuryn Anti-airlock Drainage System was attached to the Foley catheter and urine output / drainage line pressure was monitored.
Control- Critcore Urine Output MonitorA commercially available urine output monitor (Criticore, Bard Medical) was attached to the Foley catheter and urine output / drainage line pressure was monitored.
  • Accuryn (Device): Accuryn is a novel electronic urine output monitor

+Outcome measures

Primary outcomes

  • Urine Output of SOC Device Versus Accuryn [2 - 5 days]: Urine output as recorded by the SOC when connected to a standard Foley catheter and standard urinary drainage tube will be compared to urine output recorded by Accuryn and the SOC when connected to a standard Foley catheter and the Accuryn Drainage Tube.

+Eligibility

Sex: All, ages: 18 Years to any

Inclusion Criteria:

* Patient must ≥ 18 years of age
* Patient has a Foley catheter and urine collection system is in place per standard clinical decision
* Estimated length of placement of the Foley is 48 hours minimum
* Burn injury ≥ 20% and ≤ 80% total body surface area (TBSA)
* Subject or subject's legally authorized representative is able to give informed consent before entering the study

Exclusion Criteria:

* Currently pregnant or breastfeeding
* Clinical signs or symptoms of a urinary tract infection (UTI)
* Clinical signs or symptoms of a vaginal infection
* Currently has bladder or urethral trauma
* Use of investigational drug/device therapy within the past 4 weeks

+Posted results

Participant flow

MilestoneTest- Accuryn Urine Output MonitorControl- Critcore Urine Output Monitor
Started1010
Completed1010
Not completed00

Outcome results

  • Urine Output of SOC Device Versus Accuryn (Primary), Number, Patients with airlocks: Test- Accuryn Urine Output Monitor 0 (0 to 0); Control- Critcore Urine Output Monitor 5 (5 to 5)

Adverse events

ArmDeathsSerious adverse eventsOther adverse events
Test- Accuryn Urine Output Monitor000
Control- Critcore Urine Output Monitor000

+Locations (2)

FacilityCityCountryStatus
Joseph M. Still Burn Center at Doctors HospitalAugusta, GeorgiaUnited States
The University of Texas Medical Branch (UTMBP)Galveston, TexasUnited States

+Investigators